Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2024

Ultrasound-guided deployment of ProGlide™ device in transfemoral transcatheter aortic valve implantation and risk reduction of vascular complications: A propensity-matched cohort study.

Ross Daniel S, Nogic Jason, Cong Hey, Nuzzi Vincenzo, Thanikachalam Prakash, Veenuraju Shree, Gruner-Hegge Nicolai, Anwar Majid, O'Sullivan Michael, Costopoulos Charis, Davies Willaim R, Costanzo Pierluigi

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

ProGlide is a percutaneous suture-mediated closure device used in arterial and venous closure following percutaneous intervention. Risk of vascular complications from use, particularly related to failure in hemostasis, or acute vessel closure, remains significant and often related to improper suture deployment. We describe a technique of ultrasound-guided ProGlide deployment in transfemoral transcatheter aortic valve implantation (TF-TAVI).

AIMS: The aim of this study is to assess vascular outcomes for ultrasound-guided deployment of ProGlide vascular closure devices in patients undergoing TF-TAVI.

METHODS

We collected relevant clinical data of patients undergoing TAVI in a large volume centre.

PRIMARY OUTCOME: main access Valve Academic Research Consortium 3 (VARC-3) major vascular complication.

SECONDARY OUTCOME: any major/minor VARC-3 vascular complication, its type (bleed or ischemia), and treatment required (medical, percutaneous, or surgical). We performed inverse weighting propensity score analysis to compare the population undergoing ultrasound-guided versus conventional ProGlide deployment for main TAVI access. Ultrasound technique for ProGlide insertion was performed as described below.

RESULTS

Five hundred and seventeen patients undergoing TF-TAVI were included.

PRIMARY OUTCOME: In 126 (ultrasound-guided) and 391 (conventional ProGlide insertion), 0% versus 1.8% (p < 0.001) had a major VARC-3 vascular complication, respectively.

SECONDARY OUTCOME: 0.8% (one minor VARC-3 bleed) vs 4.1% (13 bleeds and three occlusions) had any VARC-3 vascular complication (major and minor) (p < 0.001). Surgical treatment of vascular complication was required in 0.8% versus 1.3% (p = NS).

CONCLUSIONS

Ultrasound-guided deployment of ProGlide for vascular closure reduced the risk of major vascular complications in a large population undergoing TAVI.

Для цитирования: Ross Daniel S, Nogic Jason, Cong Hey, Nuzzi Vincenzo, Thanikachalam Prakash, Veenuraju Shree, Gruner-Hegge Nicolai, Anwar Majid, O'Sullivan Michael, Costopoulos Charis, Davies Willaim R, Costanzo Pierluigi. Ultrasound-guided deployment of ProGlide™ device in transfemoral transcatheter aortic valve implantation and risk reduction of vascular complications: A propensity-matched cohort study.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2024. DOI: 10.1002/ccd.30999.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.