DOI 10.1364/oe.17.014180

Медицинская библиотека · Optics express · 2009

Validation of a Hartmann-Moiré wavefront sensor with large dynamic range.

Wei Xin, Van Heugten Tony, Thibos Larry

Оценочное исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

UNLABELLED: Our goal was to validate the accuracy, repeatability, sensitivity, and dynamic range of a Hartmann-Moiré (HM) wavefront sensor (PixelOptics, Inc.) designed for ophthalmic applications.

METHODS

Testing apparatus injected a 4 mm diameter monochromatic (532 nm) beam of light into the wavefront sensor for measurement. Controlled amounts of defocus and astigmatism were introduced into the beam with calibrated spherical (-20D to + 18D) and cylindrical (-8D to + 8D) lenses. Repeatability was assessed with three repeated measurements within a 2-minute period.

RESULTS

Correlation coefficients between mean wavefront measurements (n = 3) and expected wavefront vergence for both sphere and cylinder lenses were >0.999. For spherical lenses, the sensor was accurate to within 0.1D over the range from -20D to + 18D. For cylindrical lenses, the sensor was accurate to within 0.1D over the range from -8D to + 8D. The primary limitation to demonstrating an even larger dynamic range was the increasingly critical requirements for optical alignment. Sensitivity to small changes of vergence was constant over the instrument's full dynamic range. Repeatability of measurements for fixed condition was within 0.01D.

CONCLUSION

The Hartmann-Moiré wavefront sensor measures defocus and astigmatism accurately and repeatedly with good sensitivity over a large dynamic range required for ophthalmic applications.

Для цитирования: Wei Xin, Van Heugten Tony, Thibos Larry. Validation of a Hartmann-Moiré wavefront sensor with large dynamic range.. Optics express. 2009. DOI: 10.1364/oe.17.014180.