MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jocn.2010.11.012

Медицинская библиотека · Journal of clinical neuroscience : official journal of the Neurosurgical Society of Australasia · 2011

Twiddler's syndrome in a patient with a deep brain stimulation device for generalized dystonia.

Astradsson Arnar, Schweder Patrick M, Joint Carole, Green Alexander L, Aziz Tipu Z

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Deep brain stimulation (DBS) is the technique of neurostimulation of deep brain structures for the treatment of conditions such as essential tremor, dystonia, Parkinson's disease and chronic pain syndromes. The procedure uses implanted deep brain stimulation electrodes connected to extension leads and an implantable pulse generator (IPG). Hardware failure related to the DBS procedure is not infrequent, and includes electrode migration and disconnection. We describe a patient who received bilateral globus pallidus internus DBS for dystonia with initially good clinical response, but the device eventually failed. Radiographs showed multiple twisting of the extension leads with disconnection from the brain electrodes and a diagnosis of Twiddler's syndrome was made. Twiddler's syndrome was first described in patients with cardiac pacemakers. Patients with mental disability, elderly and obese patients are at increased risk. Twiddler's syndrome should be suspected whenever there is a failure of the DBS device to relieve symptoms previously responsive to stimulation. Surgical correction is usually required.

Для цитирования: Astradsson Arnar, Schweder Patrick M, Joint Carole, Green Alexander L, Aziz Tipu Z. Twiddler's syndrome in a patient with a deep brain stimulation device for generalized dystonia.. Journal of clinical neuroscience : official journal of the Neurosurgical Society of Australasia. 2011. DOI: 10.1016/j.jocn.2010.11.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.