Effects of high-profile crossover feet on gait biomechanics in 2 individuals with Syme amputation.

СтатьяРандомизированное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Slater Conrad, Hafner Brian J, Morgan Sara J

Аннотация

BACKGROUND: Prosthetic treatment options for people with ankle disarticulation (i.e., Syme amputation) are limited. Prosthetic feet designed for people with Syme amputation are often low profile to accommodate build-height restrictions, resulting in decreased energy return during gait. High-profile crossover feet that attach to the posterior proximal aspect of the prosthetic socket can bypass these restrictions and may promote a more physiologic gait pattern.

OBJECTIVES: To compare level-ground gait biomechanics and patient-reported outcomes between crossover and traditional energy-storing feet in people with Syme amputation.

STUDY DESIGN: Within-participant pilot study.

METHODS: Both participants were fit with energy-storing and crossover feet and were randomized to the order they used the feet. Participants used each foot for 2 weeks before assessment. Step length symmetry, prosthetic ankle range of motion, prosthetic-side energy return, and peak sound-side loading were determined from motion capture data obtained in a laboratory. Mobility and balance confidence were measured using standardized patient-reported outcome measures. Foot preference was assessed with an ad hoc survey.

RESULTS: Two participants with Syme amputations completed the study. Prosthetic ankle peak dorsiflexion and push-off power increased with the crossover foot compared with the energy-storing foot for both participants. Both participants reported an overall preference of the crossover foot. Changes in patient-reported outcomes did not exceed published minimum detectable change values.

CONCLUSION: Crossover feet increased prosthetic ankle range of motion and energy return compared with traditional energy-storing feet in this pilot investigation of 2 participants. Crossover feet seem to promote physiologic gait and may be a promising alternative to traditional low-profile feet for people with Syme amputation.