Paediatric anaesthesia · 2012
A randomized equivalence trial comparing the i-gel and laryngeal mask airway Supreme in children.
Сравнительное исследованиеСтатьяРандомизированное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Jagannathan Narasimhan, Sommers Katherine, Sohn Lisa E, Sawardekar Amod, Shah Ravi D, Mukherji Isabella I, Miller Steven, Voronov Polina, Seraphin Sally
Аннотация
BACKGROUND: The laryngeal mask airway Supreme (Supreme) is a new single-use supraglottic device with gastric access capability now available in all sizes for children.
OBJECTIVES: To compare the i-gel with the Supreme in children for routine airway maintenance.
MATERIALS/METHODS: One hundred and seventy children, aged 3 months to 11 years, 5-50 kg in weight, were randomly assigned to receive either the i-gel or the Supreme. The primary outcome measured was airway leak pressure. Secondary outcomes included the following: ease and time for insertion, insertion success rate, fiberoptic grade of view, ease of gastric tube placement, number of airway manipulations, quality of airway during anesthetic maintenance, and complications.
RESULTS: A total of 168 patients were assessed for the outcomes. The median (IQR [range]) airway leak pressure for the i-gel was higher than with the Supreme, 20 (18-25 [9-40]) cm H(2)O vs 17 (14-22 [10-40]) cm H(2)O, respectively (P = 0.001). There were no differences in the time for device insertion, fiberoptic grade of view, quality of airway, and complications. Median (IQR[range]) time of successful insertion of a gastric tube was faster with the Supreme, 12 (9.2-14.3 [5.2-44.2]) s than with the i-gel, 14 (11.9-19 [6.9-75]) s; P = 0.01. The number of airway manipulations during placement was higher with the i-gel than with the laryngeal mask airway Supreme (12 vs 13 patients), P = 0.02.
CONCLUSIONS: In infants and children, when a single-use supraglottic device with gastric access capabilities is required, the i-gel demonstrated higher airway leak pressures and can be a useful alternative to the Supreme.
OBJECTIVES: To compare the i-gel with the Supreme in children for routine airway maintenance.
MATERIALS/METHODS: One hundred and seventy children, aged 3 months to 11 years, 5-50 kg in weight, were randomly assigned to receive either the i-gel or the Supreme. The primary outcome measured was airway leak pressure. Secondary outcomes included the following: ease and time for insertion, insertion success rate, fiberoptic grade of view, ease of gastric tube placement, number of airway manipulations, quality of airway during anesthetic maintenance, and complications.
RESULTS: A total of 168 patients were assessed for the outcomes. The median (IQR [range]) airway leak pressure for the i-gel was higher than with the Supreme, 20 (18-25 [9-40]) cm H(2)O vs 17 (14-22 [10-40]) cm H(2)O, respectively (P = 0.001). There were no differences in the time for device insertion, fiberoptic grade of view, quality of airway, and complications. Median (IQR[range]) time of successful insertion of a gastric tube was faster with the Supreme, 12 (9.2-14.3 [5.2-44.2]) s than with the i-gel, 14 (11.9-19 [6.9-75]) s; P = 0.01. The number of airway manipulations during placement was higher with the i-gel than with the laryngeal mask airway Supreme (12 vs 13 patients), P = 0.02.
CONCLUSIONS: In infants and children, when a single-use supraglottic device with gastric access capabilities is required, the i-gel demonstrated higher airway leak pressures and can be a useful alternative to the Supreme.