DOI 10.1016/j.hrthm.2021.10.019
Predictors of perforation during lead extraction: Results of the Canadian Lead ExtrAction Risk (CLEAR) study.
Аннотация
BACKGROUND
Transvenous lead extraction can have serious adverse events, such as cardiac or vascular perforation. Risk factors have not been well characterized.
OBJECTIVE
The purpose of this study was to identify factors associated with perforation and death, and to characterize lead extraction in a large contemporary population.
METHODS
We performed a retrospective multicenter study examining patients undergoing lead extraction at 8 Canadian institutions from 1996 through 2016. Demographic and clinical data were used to identify variables associated with perforation and mortality using logistic regression modeling.
RESULTS
A total of 2325 consecutive patients (age 61.9 ±16.5 years) underwent extraction of 4527 leads. Perforation rate was 2.7% (63/2325) and 30-day mortality was 1.6% (38/2325), with mortality of 0.4% due to perforation (10/2325). Variables associated with perforation included no previous cardiac surgery (odds ratio [OR] 3.33; 95% confidence interval [CI] 1.54-7.19; P = .002), female sex (OR 3.27; 95% CI 1.91-5.60; P <.001); left ventricular ejection fraction ≥40% (OR 2.81; 95% CI 1.28-6.14; P = .010); lead age >8 years (OR 2.64; 95% CI 1.52-4.60; P <.001); ≥2 leads extracted (OR 2.49; 95% CI 1.23-5.04; P = .011); and diabetes (OR 2.12; 95% CI 1.16-3.86; P = .014). Variables associated with death included infection as indication for extraction (OR 3.85; 95% CI 1.38-10.73; P = .010); anemia (OR 3.14; 95% CI 1.38-6.61; P = .003), and patient age (OR 1.04; 95% CI 1.01-1.07; P = .012).
CONCLUSION
Risk factors associated with perforation in lead extraction include no history of cardiac surgery, female sex, preserved left ventricular ejection fraction, lead age >8 years, ≥2 leads extracted, and diabetes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.