DOI 10.1097/mpg.0000000000002666
Relative Accuracy of Bioelectrical Impedance Analysis for Assessing Body Composition in Children With Severe Obesity.
Аннотация
OBJECTIVES
The accuracy of different bioelectrical impedance analysis (BIA) devices for assessing body composition in children with obesity is unclear. We determined the relative accuracy of 2 BIA devices compared to dual x-ray absorptiometry (DXA) in obese and severely obese children.
METHODS
We measured body composition in a cross-sectional study of 78 obese children by a handheld single frequency tetrapolar BIA device (Omron), a stationary multifrequency octopolar BIA device (InBody 370) and DXA. Intermethod agreement was assessed by intraclass correlations, paired t tests, and Bland-Altman analyses.
RESULTS
Participants (37% female, age 14.8 ± 2.7 years) had mean (±standard deviation) body mass index of 36.7 ± 7.5 kg/m, body fat percentage of 46.4% ± 5.2%, and appendicular lean mass of 22.5 ± 6.0 kg by DXA. Intraclass correlations with DXA for body fat percentage were 0.39 and 0.87 for single frequency tetrapolar and multifrequency octopolar BIA devices, respectively. The single frequency tetrapolar BIA underestimated body fat percentage by 5.5% ± 2.9% (P < 0.0001). Differences between the multifrequency octopolar BIA and DXA for body fat percentage (-1.1% ± 2.8%) and appendicular lean mass (-0.3 ± 1.4 kg) were small, and 95% limits of agreement were approximately ±5%.
CONCLUSIONS
BIA machines vary in relative accuracy in measuring body composition in children who are obese and severely obese. The multifrequency octopolar BIA device accurately estimated body fat percentage and appendicular lean mass relative to DXA and has the advantage of point of care performance.