DOI 10.1097/ijg.0000000000001765

Медицинская библиотека · Journal of glaucoma · 2021

Comparative Evaluation of Rebound and Perkins Tonometers in Pediatric Glaucoma With Varied Corneal Characteristics.

Angmo Dewang, Ramesh Priyanka, Mahalingam Karthikeyan, Azmira Kishan, Pandey Shivam, Gupta Viney, Sihota Ramanjit, Dada Tanuj

СтатьяПеревод в очереди

Аннотация

PRECIS: Icare tonometer overestimated intraocular pressure (IOP) as compared with Perkins and this variation was higher in IOP>19 mm Hg and corneal opacity in patients with pediatric glaucoma.

PURPOSE: To compare the IOP measured by Icare ic200 with Perkins tonometer in pediatric glaucoma with different corneal characteristics.

METHODS

Patients of pediatric glaucoma posted for routine examination under anesthesia, age below 12 years were enrolled. All patients underwent IOP measurement with Perkins and Icare ic200 tonometer by the same observer. Basic demographic data and other relevant clinical data were recorded. Central corneal thickness (CCT), horizontal corneal diameter, and corneal characteristics such as cornea clarity was recorded.

RESULTS

A total of 194 eyes of 105 patients were analyzed. The difference between Perkins and Icare IOP was -0.816 mm Hg with the Bland-Altman plot 95% limits of agreement (LoA) from -11.194 to 9.562 mm Hg and 5.1% (10) values lying outside LoA. At IOP <19 mm Hg, the difference was -0.65 mm Hg and IOP ≥19 mm Hg, the difference was higher, -1.12 mm Hg. In the clear cornea group (123 eyes), the difference in IOP by 2 tonometers was -0.776 mm Hg with the Bland-Altman plot 95% LoA between -10.679 and 9.128 mm Hg. In hazy corneas (36 eyes), the difference in IOP was 0.531 mm Hg. The Bland-Altman plot showed 95% LoA between -6.242 and 7.303 mm Hg. In the scarred cornea group (35 eyes), the difference in IOP between the 2 was -2.343 mm Hg and the Bland-Altman plot showed wide 95% LoA from -16.302 to 11.616 mm Hg.

CONCLUSION

Icare tonometer overestimated IOP as compared with Perkins and this variation was higher in eyes with IOP≥19 mm Hg, CCT >615 μm, and scarred corneas. A moderate correlation between IOP and CCT for both tonometers was noted.

Для цитирования: Angmo Dewang, Ramesh Priyanka, Mahalingam Karthikeyan, Azmira Kishan, Pandey Shivam, Gupta Viney, Sihota Ramanjit, Dada Tanuj. Comparative Evaluation of Rebound and Perkins Tonometers in Pediatric Glaucoma With Varied Corneal Characteristics.. Journal of glaucoma. 2021. DOI: 10.1097/ijg.0000000000001765.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.