DOI 10.3928/1081597x-20150820-06
Opaque Bubble Layer Risk Factors in Femtosecond Laser-assisted LASIK.
Аннотация
PURPOSE: To determine the characteristics and risk factors for occurrence of opaque bubble layer (OBL) during femtosecond laser-assisted flap creation for LASIK.
METHODS
One hundred ninety-eight eyes of 102 consecutive patients who underwent LASIK flap creation performed with the Alcon WaveLight FS200 laser (Alcon Laboratories, Inc., Fort Worth, TX) were retrospectively analyzed in a cohort study. Preoperative manifest refraction, corneal keratometry, central corneal thickness, white-to-white corneal diameter, corneal hysteresis, corneal resistance factor, and programmed flaps parameters were collected. Digital images automatically recorded after flap creation were analyzed to measure OBL areas. Correlation tests were performed between preoperative corneal parameters and OBL areas.
RESULTS
The incidence rate of OBL was 48% (103 eyes). The mean OBL area as a percentage of the corneal flap area in the OBL group was 4.25% ± 7.16% (range: 0% to 32.9%). The central corneal thickness, corneal resistance factor, and corneal hysteresis were significantly positively correlated with the OBL area (r = 0.242, P = .001; r = 0.254, P = .028; and r = 0.351, P < .0001, respectively). Corneal hysteresis and OBL area were positively correlated, independently of the central corneal thickness and other confounder factors with standardized coefficient (r = 0.353 ± 0.227, P = .002).
CONCLUSIONS
This study confirms the already known OBL risk factors with a larger cohort and suggests for the first time that an elevated corneal hysteresis is an independent predictive risk for OBL occurrence.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.