In vitro hemodynamic testing of Amplatzer plugs for paravalvular leak occlusion after transcatheter aortic valve implantation.

Burriesci Gaetano, Peruzzo Paolo, Susin Francesca Maria, Tarantini Giuseppe, Colli Andrea

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

We aimed to in-vitro test Amplatzer devices (Amplatzer Vascular Plug II and Amplatzer Vascular Plug III, SJM, St. Paul, MN) in closing PVL generated by transcatheter balloon expandable aortic valve prosthesis in order to quantify the effective treatment of PVL.

BACKGROUND

Transcatheter aortic valve implantation (TAVI) procedures represent the treatment of choice for high risk patients. Despite evolving technologies paravalvular leak (PVL) is still a major unaddressed issue. This severe complication significantly impairs long-term survival. Percutaneous treatment of this complication is usually performed with the implantation of not specifically designed and not approved vascular devices.

METHODS

A 26 mm Sapien XT (Edwards Lifesciences, Irvine, CA) was implanted in a rubber aortic root and a semi-elliptical shape PVL was created. The vascular occluder devices were implanted in the PVL and hemodynamic performance was tested in a pulse duplicator according to international standard ISO 5840-3:2013. Different type of comparison tests together with high speed camera recording allowed us to define the global efficiency of the occluders and their interaction with the transcatheter prosthesis.

RESULTS

The results revealed that the use of vascular plugs was not per se sufficient to produce an effective or substantial reduction of PVL with a maximum efficiency of less than 50%. Recorded video showed clearly that the vascular plug always interfered with the leaflet of the prosthetic valve.

CONCLUSIONS

Currently used devices do not guarantee effective treatment of PVL and may otherwise compromise the structural integrity of the prosthetic valve implanted. Specifically designed devices are required.

CONDENSED ABSTRACT: Despite evolving technologies, paravalvular leak (PVL) is still a major unaddressed issue after transcatheter aortic valve implantation. Percutaneous treatment of this complication is usually performed with the implantation of Amplatzer devices not specifically designed and not approved for this specific use. We tested Amplatzer devices in a pulse duplicator to occlude PVL generated after implantation of a 26 mm SAPIENT XT prosthesis. The results revealed that the use of vascular plugs was not per se sufficient to produce an effective or substantial reduction of PVL. The video showed clearly that the vascular plug always interfered with the leaflet of the prosthetic valve.

Для цитирования: Burriesci Gaetano, Peruzzo Paolo, Susin Francesca Maria, Tarantini Giuseppe, Colli Andrea. In vitro hemodynamic testing of Amplatzer plugs for paravalvular leak occlusion after transcatheter aortic valve implantation.. International journal of cardiology. 2015. DOI: 10.1016/j.ijcard.2015.11.106.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.