Validating optimal function of the closed loop stimulation sensor with high right septal ventricular electrode placement in 'ablate and pace' patients.

Silberbauer John, Hong Paul S G, Veasey Rick A, Maddekar Nadeem A, Taggu Wasing, Patel Nikhil R, Lloyd Guy W, Sulke Neil

СтатьяРандомизированное исследованиеВалидационное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: The study aim was to validate the closed loop stimulation (CLS) vs. accelerometer (ACC) rate-responsive sensors with electrodes placed in the right ventricular high septal (RVHS) or right ventricular apical (RVA) lead positions in patients following 'ablate and pace' therapy for persistent atrial fibrillation.

METHODS

'Ablate and pace' patients were randomised to either RVHS or RVA electrode placement with a dual sensor device. A double-blind crossover study comparing CLS vs. ACC rate-response pacing modes was undertaken. Subjects undertook cardiopulmonary testing with constant workload light exercise followed by a ramp protocol in addition to activity of daily living assessments.

RESULTS

Twenty subjects (14 male; age, 74 +/- 8 years) were studied. Heart rate increase was greater from lying to sitting with ACC. With mental stress, heart rate increase was greater with CLS. Peak heart rates were similar for stair ascent and descent in ACC mode. With CLS mode, however, the peak heart rate was significantly lower for stair descent. There was no difference between modes in mean response time, oxygen deficit, peak VO(2), VO(2) at anaerobic threshold, peak heart rate, total exercise time and total workload. CLS function was equally optimal at both electrode sites.

CONCLUSIONS

CLS rate adaptive pacing is appropriate for 'ablate and pace' patients, and this sensor functions equally well using RVA or RVHS lead positions.

Для цитирования: Silberbauer John, Hong Paul S G, Veasey Rick A, Maddekar Nadeem A, Taggu Wasing, Patel Nikhil R, Lloyd Guy W, Sulke Neil. Validating optimal function of the closed loop stimulation sensor with high right septal ventricular electrode placement in 'ablate and pace' patients.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2009. DOI: 10.1007/s10840-009-9426-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.