Investigation of an EPID-based dose verification system for patient-specific quality assurance in a harmony device.

Yoo Dohyeon, Lee Hyunggun, Lee Sangmin, Han Soorim, Kim Taeho, Kim Changhwan, Han Min Cheol, Kim Jin Sung

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Precise patient-specific quality assurance (PSQA) is critically important in advanced treatments.

PURPOSE: This study aimed to evaluate the performance of the RadCalc software's electronic portal imaging device (EPID)-based patient specific QA function, focusing on its ability to accurately reconstruct doses from EPID images acquired on the Elekta Harmony Pro linear accelerators (LINACs).

METHODS

Beam modeling was conducted for 6-MV and 6-MV flattening filter-free (FFF) photon beams on the recently released Harmony Pro LINAC system in RadCalc. Volumetric modulated arc therapy (VMAT) plans for 6-MV (19 patients) and 6-MV FFF (10 patients) were generated using the ArcCHECK phantom to replicate clinical treatments. Dose calculations from RadCalc, using portal images, were compared to those generated by the treatment planning system (TPS) through gamma analysis at 1%/1, 2%/2, and 3%/3 mm, with a 90% minimum passing rate.

RESULTS

For 6-MV and 6-MV FFF beams, gamma analysis comparing EPID data with RadCalc showed a 98.83% passing rate at 2%/2 mm. For VMAT plans, 6-MV beams achieved the highest passing rate (99.88%) for the breast case at 3%/3 mm and the lowest (82.33%) for the lymphatic node case. Similarly, 6-MV FFF beams yielded the highest passing rate (99.87%) for the abdominal case at 3%/3 mm and the lowest (94.07%) for the bone case.

CONCLUSIONS

RadCalc showed high accuracy for dose evaluation on the Elekta Harmony Pro LINAC. Beam modeling achieved a gamma passing rate of at least 98% (2%/2 mm), demonstrating reliable performance even under minor setup variations.

Для цитирования: Yoo Dohyeon, Lee Hyunggun, Lee Sangmin, Han Soorim, Kim Taeho, Kim Changhwan, Han Min Cheol, Kim Jin Sung. Investigation of an EPID-based dose verification system for patient-specific quality assurance in a harmony device.. Journal of applied clinical medical physics. 2025. DOI: 10.1002/acm2.70201.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.