DOI 10.1016/j.joms.2010.03.015

Медицинская библиотека · Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons · 2010

Optimal degree of mouth opening for laryngeal mask airway function during oral surgery.

Sanuki Takuro, Sugioka Shingo, Hirokane Motoko, Son Hiroki, Uda Rumiko, Akatsuka Masafumi, Kotani Junichiro

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This study was performed to determine the optimal degree of mouth opening in anesthetized patients requiring laryngeal mask airway (LMA) during oral surgery.

PATIENTS AND

METHODS

A single, experienced LMA user inserted the LMA in 15 patients who were scheduled for elective oral surgery. Oropharyngeal leak pressure, intracuff pressure, and fiberoptic assessment of the LMA position were sequentially documented in 5 mouth conditions-opening of 1.4 (neutral position), 2, 3, 4, and 5 cm-and any resulting ventilatory difficulties were recorded.

RESULTS

Oropharyngeal leak pressure with the mouth open 4 cm (21.8 ± 3.2 cm H(2)O, P = .025) and 5 cm (27.3 ± 7.2 cm H(2)O, P < .001) was significantly higher than in the neutral position (18.1 ± 1.5 cm H(2)O), as was intracuff pressure (neutral position, 60.0 ± 0 cm H(2)O; 4 cm, 72.6 ± 5.1 cm H(2)O [P < .001]; and 5 cm, 86.9 ± 14.4 cm H(2)O [P < .001]). LMA position, observed by fiberoptic bronchoscopy, was unchanged by mouth opening, being similar in the 5 mouth conditions (P = .999). In addition, ventilation difficulties (abnormal capnograph curves or inadequate tidal volume) occurred in 2 of 15 patients (13%) and 7 of 15 patients (53%) (P < .001) with the mouth opening of 4 and 5 cm, respectively.

CONCLUSIONS

This study showed that a mouth opening over 4 cm led to substantial increases in oropharyngeal leak pressure and intracuff pressure of the LMA, warranting caution, because gastric insufflation, sore throat, and ventilation difficulties may occur. A mouth opening of 3 cm achieves acceptable airway conditions for anesthetized patients requiring LMA.

Для цитирования: Sanuki Takuro, Sugioka Shingo, Hirokane Motoko, Son Hiroki, Uda Rumiko, Akatsuka Masafumi, Kotani Junichiro. Optimal degree of mouth opening for laryngeal mask airway function during oral surgery.. Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons. 2010. DOI: 10.1016/j.joms.2010.03.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.