MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/15330338211060170

Медицинская библиотека · Technology in cancer research & treatment · 2021

Dosimetric Comparison of Helical Tomotherapy and Intensity-Modulated Proton Therapy in Hippocampus- and Scalp-Sparing Whole Brain Radiotherapy.

Takaoka Taiki, Tomita Natsuo, Mizuno Tomoki, Hashimoto Shingo, Tsuchiya Takahiro, Tomida Masashi, Yanagi Takeshi

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Objective

Cognitive decline and alopecia after radiotherapy are challenging problems. We aimed to compare whole brain radiotherapy (WBRT) plans reducing radiation dose to the hippocampus and scalp between helical tomotherapy (HT) and intensity-modulated proton therapy (IMPT).

Methods

We conducted a planning study of WBRT for 10 patients. The clinical target volume was defined as the whole brain excluding the hippocampus avoidance (HA) region. The prescribed dose was 30 Gy in 10 fractions to cover 95% of the target. Constraint goals were defined for the target and organs at risk (OAR).

Results

Both techniques met the dose constraints for the target and OAR. However, the coverage of the target (dose covering 95% [D95%] and 98% [D98%] of the volume) were better in IMPT than HT (HT vs IMPT: D95%, 29.9 Gy vs 30.0 Gy, P < .001; D98%, 26.7 Gy vs 28.1 Gy, P = .002). The homogeneity and conformity of the target were also better in IMPT than HT (HT vs IMPT: homogeneity index, 1.50 vs 1.28, P < .001; conformity index, 1.30 vs 1.14, P < .001). IMPT reduced the D100% of the hippocampus by 59% (HT vs IMPT: 9.3 Gy vs 3.8 Gy, P < .001) and reduced the Dmean of the hippocampus by 37% (HT vs IMPT: 11.1 Gy vs 7.0 Gy, P < .001) compared with HT. The scalp IMPT reduced the percentage of the volume receiving at least 20 Gy (V20Gy) and V10Gy compared with HT (HT vs IMPT: V20Gy, 56.7% vs 6.6%, P < .001; V10Gy, 90.5% vs 37.1%, P < .001).

Conclusion

Both techniques provided acceptable target dose coverage. Especially, IMPT achieved excellent hippocampus- and scalp-sparing. HA-WBRT using IMPT is a promising treatment to prevent cognitive decline and alopecia.

Для цитирования: Takaoka Taiki, Tomita Natsuo, Mizuno Tomoki, Hashimoto Shingo, Tsuchiya Takahiro, Tomida Masashi, Yanagi Takeshi. Dosimetric Comparison of Helical Tomotherapy and Intensity-Modulated Proton Therapy in Hippocampus- and Scalp-Sparing Whole Brain Radiotherapy.. Technology in cancer research & treatment. 2021. DOI: 10.1177/15330338211060170.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.