Hemodynamic Bedside Monitoring Instrument with Pressure and Optical Sensors: Validation and Modality Comparison.

Kaisti Matti, Panula Tuukka, Sirkiä Jukka-Pekka, Pänkäälä Mikko, Koivisto Tero, Niiranen Teemu, Kantola Ilkka

СтатьяСравнительное исследованиеВалидационное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Results from two independent clinical validation studies for measuring hemodynamics at the patient's bedside using a compact finger probe are reported. Technology comprises a barometric pressure sensor, and in one implementation, additionally, an optical sensor for photoplethysmography (PPG) is developed, which can be used to measure blood pressure and analyze rhythm, including the continuous detection of atrial fibrillation. The capabilities of the technology are shown in several form factors, including a miniaturized version resembling a common pulse oximeter to which the technology could be integrated in. Several main results are presented: i) the miniature finger probe meets the accuracy requirements of non-invasive blood pressure instrument validation standard, ii) atrial fibrillation can be detected during the blood pressure measurement and in a continuous recording, iii) a unique comparison between optical and pressure sensing mechanisms is provided, which shows that the origin of both modalities can be explained using a pressure-volume model and that recordings are close to identical between the sensors. The benefits and limitations of both modalities in hemodynamic monitoring are further discussed.

Для цитирования: Kaisti Matti, Panula Tuukka, Sirkiä Jukka-Pekka, Pänkäälä Mikko, Koivisto Tero, Niiranen Teemu, Kantola Ilkka. Hemodynamic Bedside Monitoring Instrument with Pressure and Optical Sensors: Validation and Modality Comparison.. Advanced science (Weinheim, Baden-Wurttemberg, Germany). 2024. DOI: 10.1002/advs.202307718.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.