Clinical chemistry and laboratory medicine · 2023

Multicenter study to compare the diagnostic performance of CLIA vs. FEIA transglutaminase IgA assays for the diagnosis of celiac disease.

Многоцентровое исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Castelijn Daan A R, Mulder A H Leontine, van der Pol Pieter, Hollander Jolien C, Kuiper Tietie, Bijnens Caroline, Damoiseaux Jan, Bontkes Hetty J

Аннотация

OBJECTIVES: Celiac disease (CD) is an immune-mediated enteropathy driven by gluten intake. Presence of tTG-IgA antibodies is important for the diagnosis. However, different tTG-IgA assays are used and test performance may vary. Therefore, a retrospective multicenter study was performed to compare the diagnostic performance of three assays.

METHODS: The fluorescence enzyme-linked immunoassay (FEIA) EliA Celikey IgA (Phadia), the chemiluminescence immunoassays (CLIA) h-tTG IgA QUANTA Flash® (Inova Diagnostics) and the anti-tTG ChLIA IgA (Euroimmun) were compared. Diagnostic samples from CD cases (95 adults; 65 children) and controls (479 adults; 253 children) were included. Samples were blinded and reanalyzed on all platforms.

RESULTS: A high quantitative correlation between platforms was found (p<0.0001). Both CLIA were more sensitive (adults 100%; children 100%) compared to the FEIA (adults 88.4%; children 96.6%). Specificity of all assays was high (≥97.6%) with the FEIA having the highest specificity. A cut-off based on receiver operator characteristic analysis (6.5 U/mL) improved the sensitivity of the FEIA (adults 95.8%; children 100%) without affecting specificity. Cut-off values for the CLIA assays did not need further optimization. With the FEIA, 71% of pediatric cases had a tTG-IgA level ≥10× upper limit of normal compared to 91 and 92% with QUANTA Flash and ChLIA, respectively.

CONCLUSIONS: All platforms have high diagnostic accuracy. The CLIA assays are more sensitive compared to the FEIA assay. A lower cut-off for the FEIA improves diagnostic performance, particularly in adult cases that, as demonstrated in this study, present with lower tTG-IgA levels compared to pediatric cases.