MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1118/1.596189

Медицинская библиотека · Medical physics · 1988

Energy and rate dependence of diagnostic x-ray exposure meters.

Wagner L K, Cerra F, Conway B, Fewell T R, Ohlhaber T R

СтатьяПеревод пока не подготовлен

Аннотация

Variations in x-ray exposure measurements among a variety of contemporary diagnostic exposure meters are investigated. Variations may result from systematic errors due to calibration, beam-quality dependence and exposure-rate dependence. It is concluded that the majority of general purpose diagnostic meters will agree to within 10% of each other if exposure rates are below 1.3 mC kg-1S-1 of air (5 R s-1) and beam qualities are typical for general purpose radiology, excluding mammography. For exposure rates comparable to those in barium enema radiography the variations can range up to 25% or more. Variations up to 40% were observed among general purpose exposure meters at mammographic beam qualities. In the mammographic range, mammographic (thin window) exposure meters varied by no more than 2%.

Для цитирования: Wagner L K, Cerra F, Conway B, Fewell T R, Ohlhaber T R. Energy and rate dependence of diagnostic x-ray exposure meters.. Medical physics. 1988. DOI: 10.1118/1.596189.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.