Stroke · 2007
Power M-Mode Doppler and single-gate spectral analysis using a 2-MHz pulsed-wave Doppler transducer to directly detect cervical internal carotid artery stenosis: use of the continuity principle: report of a novel technique.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Popa Vasile N, Spencer Merrill P, Lion Charlene L, Felberg Robert A
Аннотация
BACKGROUND AND PURPOSE: We hypothesized that direct cervical investigation with Power M-Mode Doppler (PMD) combined with single-gate Doppler spectral analysis (SGDSA) using a 2-MHz pulsed-wave Doppler transducer would show reasonable accuracy parameters when compared with standard color-coded carotid duplex ultrasound (CDU).
METHODS: We prospectively screened for cervical internal carotid artery (ICA) stenosis by direct observation using a 2 MHz PMD/SGDSA device. PMD identified the artery (location, depth, flow direction) and SGDSA assessed waveform; peak systolic, end diastolic, and mean flow velocities (MFV) of the common carotid artery; cervical ICA proximally and distally; and external carotid artery. Diagnostic accuracy was compared with concurrent carotid duplex ultrasound. The continuity principle was applied using the proximal/distal cervical ICA MFV ratio.
RESULTS: We examined 456 vessels (228 patients). Using ICA proximally/ICA distally MFV ratio of 1.5 or greater or absence of ICA signature, for 40% to 59% or greater stenosis (including occlusions), sensitivity was 75.4%, specificity 99.8%, positive predictive value 97.7%, negative predictive value 96.6%, and accuracy 96.7%. For MFV ratio 1.6 or greater or absence of ICA signature and 60% to 79% or greater stenosis (including occlusions), sensitivity was 92.3%, specificity 98.1%, positive predictive value 81.8%, negative predictive value 99.3%, and accuracy 97.6%.
CONCLUSIONS: Use of combined PMD and SGDSA to directly observe the extracranial ICA is reasonably accurate compared with carotid duplex ultrasound. Using the MFV ratio of proximal/distal extracranial ICA improves accuracy parameters and provides a quick and effective bedside screen for ICA stenosis. This novel technique should be considered part of the standard PMD/transcranial Doppler examination.
METHODS: We prospectively screened for cervical internal carotid artery (ICA) stenosis by direct observation using a 2 MHz PMD/SGDSA device. PMD identified the artery (location, depth, flow direction) and SGDSA assessed waveform; peak systolic, end diastolic, and mean flow velocities (MFV) of the common carotid artery; cervical ICA proximally and distally; and external carotid artery. Diagnostic accuracy was compared with concurrent carotid duplex ultrasound. The continuity principle was applied using the proximal/distal cervical ICA MFV ratio.
RESULTS: We examined 456 vessels (228 patients). Using ICA proximally/ICA distally MFV ratio of 1.5 or greater or absence of ICA signature, for 40% to 59% or greater stenosis (including occlusions), sensitivity was 75.4%, specificity 99.8%, positive predictive value 97.7%, negative predictive value 96.6%, and accuracy 96.7%. For MFV ratio 1.6 or greater or absence of ICA signature and 60% to 79% or greater stenosis (including occlusions), sensitivity was 92.3%, specificity 98.1%, positive predictive value 81.8%, negative predictive value 99.3%, and accuracy 97.6%.
CONCLUSIONS: Use of combined PMD and SGDSA to directly observe the extracranial ICA is reasonably accurate compared with carotid duplex ultrasound. Using the MFV ratio of proximal/distal extracranial ICA improves accuracy parameters and provides a quick and effective bedside screen for ICA stenosis. This novel technique should be considered part of the standard PMD/transcranial Doppler examination.