DOI 10.1097/00006123-200206000-00017
Long-term hardware-related complications of deep brain stimulation.
Аннотация
OBJECTIVE
To determine the incidence of long-term hardware-related complications of deep brain stimulation (DBS).
METHODS
The study design is a retrospective chart review of a single-surgeon, single-institution experience with DBS in 84 consecutive cases from 1993 to 1999. Only patients with a minimum follow-up of 1 year were considered. Five patients were excluded because trial stimulation failed to achieve pain relief (n = 4) or because the procedure was aborted owing to hemorrhage (n = 1). Seventy-nine patients received 124 permanent DBS electrode implants.
RESULTS
The mean follow-up period was 33 months, and the cumulative follow-up time was 217 patient-years or 310 electrode-years. Overall, 20 patients (25.3%) had 26 hardware-related complications involving 23 (18.5%) of the electrodes. There were 4 lead fractures, 4 lead migrations, 3 short or open circuits, 12 erosions and/or infections, 2 foreign body reactions, and one cerebrospinal fluid leak. The hardware-related complication rate per electrode-year was 8.4%. The most common complications were related to the electrode connectors. A significant finding was a high number of complications involving erosions or infections, which occurred in 7 of 12 instances as a late complication (beyond 12 mo).
CONCLUSION
Long-term follow-up reveals that hardware-related complications occur in a significant number of patients. Factors that lead to such complications must be identified and addressed to maximize the important benefits of DBS therapy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.