Color Doppler ultrasound for surveillance following EVAR as the primary tool.
Аннотация
OBJECTIVE
As aneurysm-related events and rupture is not eliminated, postoperative lifelong surveillance is mandatory after endovascular aneurysm repair (EVAR). For surveillance colored Doppler ultrasound (CDUS) is a standard method of noninvasive evaluation having the advantages of availability, cost-effectiveness, and lack of nephrotoxicity and radiation. We evaluated CDUS for primary surveillance tool after elective EVAR by comparing with computerized tomography.
METHODS
Between January 2018 and March 2020, 84 consecutive post-EVAR patients were evaluated. First, CDUS was performed by two Doppler operators from the Radiology Department and then computed tomographic angiography (CTA) was performed. The operators were blind to CTA reports. A reporting protocol was organized for endoleak detection and largest aneurysm diameter.
RESULTS
Among 84 patients, there were 11 detected endoleaks (13.1%) with CTA and seven of them was detected with CDUS (r = .884, p < .001). All Type I and III endoleaks were detected perfectly. There is an insufficiency in detecting low flow by CDUS. Eliminating this frailty, there was a strong correlation of aneurysm sac diameter measurement between CTA and CDUS (r = .777, p < .001). The sensitivity and specificity of CDUS was 63.6% and 100%, respectively. The accuracy was 95.2%. Positive and negative predictive values were 100% and 94.8%. Bland-Altman analysis and linear regression analysis showed no proportional bias (mean difference of 1.5 ± 2.2 mm, p = .233).
CONCLUSIONS
For surveillance, CDUS promises accurate results without missing any potential complication requiring intervention as Type I or III endoleak. Lack of detecting Type II endoleaks may be negligible as sac enlargement was the key for reintervention in this situation and CDUS has a remarkably high correlation with CTA in sac diameter measurement. CDUS may be a primary surveillance tool for EVAR and CTA will be reserved in case of aneurysm sac enlargement, detection of an endoleak, inadequate CDUS, or in case of unexplained abdominal symptomatology. By this way we not only avoid ionizing radiation and nephrotoxic agents, but also achieve cost saving issue also.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.