DOI 10.1016/s0268-0033(01)00102-4

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2002

Mechanical conditions in the internal stabilization of proximal tibial defects.

Duda Georg N, Mandruzzato Francesco, Heller Markus, Kassi Jean-Pierre, Khodadadyan Cyros, Haas Norbert P

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The goal was to design a method which would permit an assessment of the suitability of a newly developed implant under physiological-like loading conditions. Information obtained from such an analysis is expected to delineate more clearly the indications for a new device prior to clinical utilization.

DESIGN

In vitro mechanical stiffness testing and finite element analysis.

METHODS

From in vitro testing of proximal tibiae with defects, the stiffness of an internal stabilization system was determined. Using a finite element model, the loading of both the implant and bone was analyzed including all muscle forces. The variation in implant loading and interfragmentary strain for different defect locations was also investigated.

RESULTS

Conventional stiffness testing demonstrated the comparability of the experimental findings with the finite element predictions. Under physiological-like loading the implant experienced high bending and von Mises stresses if defects in the region of the shaft were stabilized. A short working length increased implant loading up to the yield strength of the material.

CONCLUSIONS

The finite element analysis illustrated the appropriateness of this new device for proximal defects of the tibia, but the implant should be used with hesitation in fractures or defects extending into the diaphyseal region of the bone.

RELEVANCE: This new analytical approach helped to identify clinical indications for the implant in which its mechanical attributes would prove advantageous.

Для цитирования: Duda Georg N, Mandruzzato Francesco, Heller Markus, Kassi Jean-Pierre, Khodadadyan Cyros, Haas Norbert P. Mechanical conditions in the internal stabilization of proximal tibial defects.. Clinical biomechanics (Bristol, Avon). 2002. DOI: 10.1016/s0268-0033(01)00102-4.