Therapeutic drug monitoring · 2013
Micromethod for quantification of cinacalcet in human plasma by liquid chromatography-tandem mass spectrometry using a stable isotope-labeled internal standard.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Cangemi Giuliana, Barco Sebastiano, Verrina Enrico E, Scurati Samuele, Melioli Giovanni, Della Casa Alberighi Ornella
Аннотация
BACKGROUND: Cinacalcet hydrochloride is a calcimimetic agent indicated for the treatment of secondary hyperparathyroidism in dialysis-dependent patients with chronic kidney disease. In the context of a pharmacokinetic (PK)/pharmacodynamic study of cinacalcet in dialysis-dependent chronic kidney disease children with secondary hyperparathyroidism, we describe the development and validation of a new, rapid, simple, and economical liquid chromatography-tandem mass spectrometry (LC-MS/MS) micromethod for quantifying cinacalcet plasma concentrations.
METHODS: Cinacalcet was analyzed in 50-μL plasma samples over a wide range of concentrations (0.1-100 ng/mL) by LC-MS/MS after protein precipitation and addition of deuterated cinacalcet as the internal standard. Cinacalcet was quantified using selective reaction monitoring of the specific transition m/z 358.1 > 155.1, with the 361.1 > 158.1 transition used for the internal standard. The suitability of the assay for clinical PK studies was evaluated using data from a pilot PK study in a pediatric patient.
RESULTS: The overall turnaround time for the assay was 20 minutes. The lower limit of quantification of the method was 0.1 ng/mL. Intraassay imprecision and inaccuracy for quality control samples ranged from 2.8% to 9% and 100% to 102%, respectively. Interassay imprecision and inaccuracy ranged from 6.9% to 8.5% and 99% to 103%, respectively. The overall recovery ranged from 90% to 106%. No ion suppression due to matrix effects was found with different preanalytical conditions, such as hemolysis, lipemia, and hyperuricemia.
CONCLUSIONS: This LC-MS/MS micromethod provides high specificity, precision, and accuracy for rapid quantification of cinacalcet plasma concentrations, and it is suitable for application in pediatric PK studies; it also has potential for use in the establishment of target ranges and ultimately routine therapeutic drug monitoring to optimize cinacalcet dosing.
METHODS: Cinacalcet was analyzed in 50-μL plasma samples over a wide range of concentrations (0.1-100 ng/mL) by LC-MS/MS after protein precipitation and addition of deuterated cinacalcet as the internal standard. Cinacalcet was quantified using selective reaction monitoring of the specific transition m/z 358.1 > 155.1, with the 361.1 > 158.1 transition used for the internal standard. The suitability of the assay for clinical PK studies was evaluated using data from a pilot PK study in a pediatric patient.
RESULTS: The overall turnaround time for the assay was 20 minutes. The lower limit of quantification of the method was 0.1 ng/mL. Intraassay imprecision and inaccuracy for quality control samples ranged from 2.8% to 9% and 100% to 102%, respectively. Interassay imprecision and inaccuracy ranged from 6.9% to 8.5% and 99% to 103%, respectively. The overall recovery ranged from 90% to 106%. No ion suppression due to matrix effects was found with different preanalytical conditions, such as hemolysis, lipemia, and hyperuricemia.
CONCLUSIONS: This LC-MS/MS micromethod provides high specificity, precision, and accuracy for rapid quantification of cinacalcet plasma concentrations, and it is suitable for application in pediatric PK studies; it also has potential for use in the establishment of target ranges and ultimately routine therapeutic drug monitoring to optimize cinacalcet dosing.
Предметные рубрики MeSH
AdultCalcimimetic AgentsChildChromatography, LiquidCinacalcetClinical Trials as TopicDrug MonitoringHumansHyperparathyroidismIsotope LabelingMicrochemistryNaphthalenesPilot ProjectsQuality ControlReference StandardsRenal DialysisRenal Insufficiency, ChronicReproducibility of ResultsTandem Mass Spectrometry