Multicenter multiplatform pattern-of-practice analysis of single-isocenter multitarget stereotactic radiosurgery.

Thomann Benedikt, Fechter Tobias, Fischer Johannes, Runz Armin, Roers Julian, Ludwig Ute, Grehn Melanie, Grohmann Maximilian, Ziemann Christian, Judge Michael, Baus Wolfgang, Grahle Michelle, Walke Matthias, Bathen Bastian, Köhn Janett, Käthner Paul, Shariff Maya, Matthis Rebecca, Fleckenstein Jens, Großmann Sascha, Streller Tino, Howitz Simon, Priegnitz Marlen, Weigel Rocco, Winkler Peter, Blanck Oliver, Schmitt Daniela, Beck Jurgen, Machein Marcia, Pappas Evangelos, Popp Ilinca, Reiner Michael, Karger Christian P, Moustakis Christos, Bock Michael, Grosu Anca-Ligia, Baltas Dimos

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Single-isocenter multitarget stereotactic radiosurgery (SIMT SRS) offers enhanced clinical efficiency for treating multiple brain metastases. However, it introduces additional uncertainties, such as off-center dose and beam profile inaccuracies, as well as quality assurance (QA) challenges, complicating its implementation. This study aims to evaluate different SIMT SRS approaches.

METHODS

We collected and analyzed SIMT SRS protocol and infrastructure parameters from 23 radiotherapy centers across Germany, Austria, and Switzerland, encompassing immobilization systems, computed tomography (CT) protocols, linear accelerators, treatment planning systems, beam configurations, imaging techniques, and QA practices. Consensus, deviations, and compliance with current guidelines were assessed. Subsequent studies will include on-site measurements, evaluation of treatment plan quality and delivery accuracy, and correlation of these findings with the analyzed protocols to identify potential links between protocol parameters and clinical outcomes.

RESULTS

There is consensus (at least 80% agreement) for a CT slice thickness of ≤ 1 mm, the need for six-degree-of-freedom patient setup correction, and noncoplanar treatment. There is notable variability for intrafraction imaging (used by 70%), minimum accepted planning target volume diameter (ranging from 2-10 mm), SRS QA, and general plan parameters, such as photon energy and number of treatment fields. There is also high variability in employed linear accelerator models and treatment planning systems.

CONCLUSION

These findings highlight a lack of standardization in SIMT SRS practices. Combined with future measurements correlating protocols to treatment quality, our study will provide a foundation for recommendations to support the safe and standardized implementation of SIMT SRS.

Для цитирования: Thomann Benedikt, Fechter Tobias, Fischer Johannes, Runz Armin, Roers Julian, Ludwig Ute, Grehn Melanie, Grohmann Maximilian, Ziemann Christian, Judge Michael, Baus Wolfgang, Grahle Michelle, Walke Matthias, Bathen Bastian, Köhn Janett, Käthner Paul, Shariff Maya, Matthis Rebecca, Fleckenstein Jens, Großmann Sascha, Streller Tino, Howitz Simon, Priegnitz Marlen, Weigel Rocco, Winkler Peter, Blanck Oliver, Schmitt Daniela, Beck Jurgen, Machein Marcia, Pappas Evangelos, Popp Ilinca, Reiner Michael, Karger Christian P, Moustakis Christos, Bock Michael, Grosu Anca-Ligia, Baltas Dimos. Multicenter multiplatform pattern-of-practice analysis of single-isocenter multitarget stereotactic radiosurgery.. Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]. 2025. DOI: 10.1007/s00066-025-02424-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.