DOI 10.1016/j.jocmr.2025.101926

Медицинская библиотека · Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2025

Time-efficient simultaneous fat and water cardiac cine imaging using spiral magnetic resonance imaging.

Chao Tzu Cheng, Wang Dinghui, Pipe James G, Leiner Tim

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiac cine imaging is routinely used in patient with suspected or known cardiac dysfunction. Water and fat (W/F) separated cardiovascular magnetic resonance (CMR) will be helpful to distinguish adipose tissue, blood, and myocardium. Inclusion of a multi-echo acquisition in the conventional balanced steady-state free precession (bSSFP) cine sequence can introduce artifacts and reduce temporal resolution. Spiral MRI is known for its signal-to-noise ratio (SNR) efficiency and has the potential to improve temporal efficiency for W/F separated cine imaging. The present work implements a spoiled gradient echo sequence (SPGR) with spiral trajectory to obtain W/F separated cine images simultaneously.

METHODS

Three different sequences were performed for comparison, a Cartesian 2-TE bSSFP sequence, a Cartesian 3-TE bSSFP sequence, and the proposed spiral SPGR sequence. Five volunteers were recruited for the scans on a 1.5T scanner with spatial resolution 1.7×1.7×8.0mm3 over a 400×400mm2 FOV. In addition to qualitative comparisons, a quantitative measurement is performed in terms of the contrast-to-noise ratio (CNR).

RESULTS

The proposed method to obtain W/F separated cine images provides better temporal efficiency and fewer artifacts compared to conventional Cartesian bSSFP sequences. The 2-TE bSSFP features the highest artifact level, including susceptibility artifacts and fat/water swaps. The proposed method reduces scan time by approximately 50% with similar spatial and temporal resolution with lower specific absorption rate (SAR). The contrast between the blood pool and myocardium is higher when using the spiral readout (p≤0.05). The results suggest that the presented sequence has potential to facilitate simultaneous imaging for water and fat components in a cine scan while shortening exam time and lowering SAR.

Для цитирования: Chao Tzu Cheng, Wang Dinghui, Pipe James G, Leiner Tim. Time-efficient simultaneous fat and water cardiac cine imaging using spiral magnetic resonance imaging.. Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance. 2025. DOI: 10.1016/j.jocmr.2025.101926.