Impact of tissue photon attenuation in small animal cardiac PET imaging.

Hayakawa Nobuyuki, Yamane Tomohiko, Arias-Loza Anahi-Paula, Shinaji Tetsuya, Wakabayashi Hiroshi, Lapa Constantin, Werner Rudolf A, Javadi Mehrbod S, Pelzer Theo, Higuchi Takahiro

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Tissue photon attenuation is one of the essential artifacts requiring correction in clinical cardiac positron emission tomography (PET) imaging. However, due to small body size its impact on diagnostic accuracy in small rodents is considered to be limited or even ignorable. The present cardiac PET study compares lean and obese rats to determine the influence of tissue attenuation on quantitative assessment as well as regional tracer distribution.

METHODS

A dedicated small animal PET system equipped with a 57Co rotating source for transmission was used. To assess the impact of tissue attenuation in rats with different body sizes, cardiac 18F-FDG -PET studies for Zucker diabetic fatty rats (obese rats) and Zucker lean rats (lean rats) were performed. The radiotracer activity reduction by attenuation was compared between the two groups. Regional tracer distribution calculated with and without attenuation correction was also assessed.

RESULTS

The chest diameter was significantly longer in obese than in lean rats (5.6±0.3cm in obese and 4.5±0.2cm in lean rats, p<0.0001). Whereas the activity reduction by attenuation was significantly greater in obese than in lean rats (44.1±2.5% and 5.1±3.1%, p<0.0001), the regional variation of tissue attenuation among the ventricular walls was minimal in both lean (p=0.73) and obese rats (p=0.65).

CONCLUSION

Attenuation correction is indispensable for accurate comparison of cardiac tracer activity between animals with different body size, whereas it can be omitted for evaluation of regional tracer distribution.

Для цитирования: Hayakawa Nobuyuki, Yamane Tomohiko, Arias-Loza Anahi-Paula, Shinaji Tetsuya, Wakabayashi Hiroshi, Lapa Constantin, Werner Rudolf A, Javadi Mehrbod S, Pelzer Theo, Higuchi Takahiro. Impact of tissue photon attenuation in small animal cardiac PET imaging.. International journal of cardiology. 2016. DOI: 10.1016/j.ijcard.2016.11.119.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.