Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2019
Head-to-head comparison of diagnostic accuracy of stress-only myocardial perfusion imaging with conventional and cadmium-zinc telluride single-photon emission computed tomography in women with suspected coronary artery disease.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mannarino Teresa, Assante Roberta, Ricciardi Carlo, Zampella Emilia, Nappi Carmela, Gaudieri Valeria, Mainolfi Ciro Gabriele, Di Vaia Eugenio, Petretta Mario, Cesarelli Mario, Cuocolo Alberto, Acampa Wanda
Аннотация
BACKGROUND: Breast attenuation may impact the diagnostic accuracy of stress myocardial perfusion imaging (MPI) with single-photon emission computed tomography (SPECT). We compared the performance of conventional (C)-SPECT and cadmium-zinc-telluride (CZT)-SPECT systems in women with low-intermediate likelihood of coronary artery disease (CAD).
METHODS AND RESULTS: A total of 109 consecutive women underwent stress-optional rest MPI by both C-SPECT and CZT-SPECT. In the overall study population, a weak albeit significant correlation between total perfusion defect (TPD) measured by C-SPECT and CZT-SPECT was observed (r = 0.38, P < .001) and at Bland-Altman analysis the mean difference in TPD (C-SPECT minus CZT-SPECT) was 2.40% (P < .001). Overall concordance of semi-quantitative diagnostic performance between C-SPECT and CZT-SPECT was observed in 52 (48%) women with a κ value of 0.09. Normalcy rate was significantly higher using CZT-SPECT compared to C-SPECT (P < .001). Machine learning analysis performed through the implementation of J48 algorithm proved that CZT-SPECT has higher sensitivity, specificity, and accuracy than C-SPECT.
CONCLUSIONS: In women with low-intermediate likelihood of CAD, there is a poor concordance of diagnostic performance between C-SPECT and CZT-SPECT, and CZT-SPECT allows better normalcy rate detection compared to C-SPECT.
METHODS AND RESULTS: A total of 109 consecutive women underwent stress-optional rest MPI by both C-SPECT and CZT-SPECT. In the overall study population, a weak albeit significant correlation between total perfusion defect (TPD) measured by C-SPECT and CZT-SPECT was observed (r = 0.38, P < .001) and at Bland-Altman analysis the mean difference in TPD (C-SPECT minus CZT-SPECT) was 2.40% (P < .001). Overall concordance of semi-quantitative diagnostic performance between C-SPECT and CZT-SPECT was observed in 52 (48%) women with a κ value of 0.09. Normalcy rate was significantly higher using CZT-SPECT compared to C-SPECT (P < .001). Machine learning analysis performed through the implementation of J48 algorithm proved that CZT-SPECT has higher sensitivity, specificity, and accuracy than C-SPECT.
CONCLUSIONS: In women with low-intermediate likelihood of CAD, there is a poor concordance of diagnostic performance between C-SPECT and CZT-SPECT, and CZT-SPECT allows better normalcy rate detection compared to C-SPECT.
Предметные рубрики MeSH
AgedCadmiumCoronary AngiographyCoronary Artery DiseaseElectrocardiographyExercise TestFemaleGamma CamerasHumansImage Processing, Computer-AssistedMachine LearningMiddle AgedMyocardial Perfusion ImagingPerfusionReproducibility of ResultsRisk FactorsSemiconductorsSoftwareTelluriumTomography, Emission-Computed, Single-PhotonZinc