OBJECTIVE

Because of concerns about the safety and environmental impact of mercury, aneroid sphygmomanometers have replaced mercury-filled devices for blood pressure (BP) measurements. Despite this change, few studies have compared BP measurements between the 2 devices.

METHODS

The SEARCH for Diabetes in Youth Study conducted a comparison of aneroid and mercury devices among 193 youth with diabetes (48% boys, aged 12.9 ± 3.7 years; 89% type 1). Statistical analyses included estimating Pearson correlation coefficients, Bland-Altman plots, paired t tests, and fitting regression models, both overall and stratified by age (<10 vs ≥10-18 years).

RESULTS

Mean mercury and aneroid systolic and diastolic BPs were highly correlated. For the entire group, there was no significant difference in mean systolic BP using the aneroid device, but there was a -1.53 ± 5.06 mm Hg difference in mean diastolic BP. When stratified by age, a lower diastolic BP (-1.78 ± 5.2 mm Hg) was seen in those ≥10 to 18 years using the aneroid device. No differences in systolic BP were observed, and there were no differences in BP by device in individuals <10 years. Regression analyses did not identify any explanatory variables.

CONCLUSIONS

Although a small discrepancy between diastolic BP measurements from aneroid versus mercury devices exists, this variation is unlikely to be clinically significant, suggesting that either device could be used in research or clinical settings.

Для цитирования: Shah Amy S, Dolan Lawrence M, D'Agostino Ralph B, Standiford Debra, Davis Cralen, Testaverde Lisa, Pihoker Catherine, Daniels Stephen R, Urbina Elaine M, SEARCH for Diabetes in Youth Study Group. Comparison of mercury and aneroid blood pressure measurements in youth.. Pediatrics. 2012. DOI: 10.1542/peds.2011-3087.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.