m+Medelevator.RU
← Поиск публикаций
Journal of clinical monitoring and computing · 2023

Are NIRS-derived cerebral autoregulation and ABPopt values different between hemispheres in hypoxic-ischemic brain injury patients following cardiac arrest?

Письмо в редакциюНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Hazenberg L, Aries Mjh, Beqiri E, Mess W H, van Mook Wnka, Delnoij T, Zeiler F A, van Kuijk Smj, Tas J

Аннотация

PURPOSE: Near-infrared spectroscopy (NIRS) has been suggested as a non-invasive monitoring technique to set cerebral autoregulation (CA) guided ABP targets (ABPopt) in comatose patients with hypoxic-ischemic brain injury (HIBI) following cardiac arrest. We aimed to determine whether NIRS-derived CA and ABPopt values differ between left and right-sided recordings in these patients.

METHODS: Bifrontal regional oxygen saturation (rSO2) was measured using INVOS or Fore-Sight devices. The Cerebral Oximetry index (COx) was determined as a CA measure. ABPopt was calculated using a published algorithm with multi-window weighted approach. A paired Wilcoxon signed rank test and intraclass correlation coefficients (ICC) were used to compare (1) systematic differences and (2) degree of agreement between left and right-sided measurements.

RESULTS: Eleven patients were monitored. In one patient there was malfunctioning of the right-sided optode and in one patient not any ABPopt value was calculated. Comparison of rSO2 and COx was possible in ten patients and ABPopt in nine patients. The average recording time was 26 (IQR, 22-42) hours. The ABPopt values were not significantly different between the bifrontal recordings (80 (95%-CI 76-84) and 82 (95%-CI 75-84) mmHg) for the left and right recordings, p = 1.0). The ICC for ABPopt was high (0.95, 0.78-0.98, p < 0.001). Similar results were obtained for rSO2 and COx.

CONCLUSION: We found no differences between left and right-sided NIRS recordings or CA estimation in comatose and ventilated HIBI patients. This suggests that in these patients without signs of localized pathology unilateral recordings might be sufficient to estimate CA status or provide ABPopt targets.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты