Automatic non-invasive blood pressure measurement in left ventricular assist device patients with a photoplethysmography assisted device.

Nagy Péter, Sax Balázs, Kozák Anna, Merkely Béla, Osztheimer István, Jobbágy Ákos

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Non-invasive measurement of blood pressure in patients with a left ventricular assist device (LVAD) is challenging due to the mechanical properties of these circulatory support devices. Keeping blood pressure in a target range is crucial to reduce the risk of LVAD complications. As current assessment methods require trained personnel resulting in infrequent monitoring, devices for home-based measurements are sorely needed.

OBJECTIVES

In this paper, we report a measurement method and a custom-made monitoring device based on photoplethysmography (PPG) enabling automated measurement of the mean arterial pressure (MAP) of patients with LVAD.

METHODS

The method and the device were tested on 21 adult patients with LVAD, and the estimated MAP values were compared to MAP values measured simultaneously by a human expert applying the commonly used Doppler-based method.

RESULTS

Results of the proposed method showed highly significant correlation with Doppler-based MAP values (R = 0.85 for inflation, p < 0.001; R = 0.96 for deflation, p<0.001). The mean difference of the proposed method to the Doppler-based MAP values was 1.48 ± 5.43 mmHg for inflation and -0.19 ± 2.71 mmHg for deflation.

CONCLUSION

The results demonstrate that the proposed method is a promising direction to achieve accurate, automated, non-invasive BP measurement, applicable in home health monitoring for patients with LVAD.

Для цитирования: Nagy Péter, Sax Balázs, Kozák Anna, Merkely Béla, Osztheimer István, Jobbágy Ákos. Automatic non-invasive blood pressure measurement in left ventricular assist device patients with a photoplethysmography assisted device.. The International journal of artificial organs. 2023. DOI: 10.1177/03913988231168159.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.