Does the Gamma Knife dose rate affect outcomes in radiosurgery for trigeminal neuralgia?
Аннотация
OBJECT: The object of this study was to determine whether the radiation dose rate affects clinical outcomes in patients who undergo stereotactic Gamma Knife surgery (GKS) to manage typical trigeminal neuralgia (TN).
METHODS
The authors retrospectively studied pain relief in 165 patients with medically intractable TN, who underwent 80-Gy GKS using a single 4-mm collimator between 1994 and 2005. No patient had received prior radiation treatment. The measured relative helmet output factor of the Gamma Knife was 0.8 throughout this interval, and the dose rate varied from 1.21 Gy/minute to 3.74 Gy/minute (median 2.06 Gy/minute). Irradiation time varied from 26.73 to 95.11 minutes. The authors divided patients into a low-dose-rate (LDR) group, in which the dose rate varied from 1.21 to 2.05 Gy/minute, and a high-dose-rate (HDR) group, in which the dose rate varied from 2.06 to 3.74 Gy/minute. Post-GKS, the patients' pain control was determined using the Barrow Neurological Institute (BNI) pain scale. There was no statistically significant difference between groups with respect to history of prior microvascular decompression (p = 0.410) or peripheral neuroablative procedures (p = 0.583). The length of symptoms in patients varied from 3 to 414 months with a median of 84 months (p = 0.698). Median follow-up was 26 months with a maximum of 139 months.
RESULTS
Initial pain relief was obtained in 71% of patients in the LDR group and 78% in the HDR group (p = 0.547). Patients who initially obtained improved pain relief (BNI Scores I-IIIa) after GKS maintained pain control for median durations of 52 months (LDR group) and 54 months (HDR group) (p = 0.403). New or increased facial sensory dysfunction was found in 14.5% of patients in the LDR group and in 19.3% of patients in the HDR group (p = 0.479).
CONCLUSIONS
The authors found that the GKS dose rate did not affect pain control or morbidity within the range of 1.21-3.74 Gy/minute. Cobalt 60 source decay did not affect outcomes of GKS for TN pain management, even for dose rates approximating a 2-half-life decay of the isotope.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.