Impact of percutaneous closure device type on vascular and bleeding complications after TAVR: A post hoc analysis from the BRAVO-3 randomized trial.
Аннотация
BACKGROUND/
OBJECTIVE
Prostar XL (PS) and ProGlide (PG) are common vascular closure devices (VCD) used in TAVR via transfemoral vascular approach. The impact of these VCD on vascular and bleeding complications remains unclear.
METHODS
The BRAVO-3 trial randomized 802 patients undergoing transfemoral TAVR. We stratified patients according to type of VCD used and examined the 30-day incidence of major or minor vascular complications, major bleeding (BARC ≥3b), AKI and major adverse cardiac and cerebrovascular events (MACCE; death, myocardial infarction or stroke).
RESULTS
A total of 746 (93%) patients were treated with either PS (n = 352, 47%) or PG (n = 394, 53%) VCD, without significant differences in successful deployment rate (PS 322 [91.2%] vs. PG 373 [94.2%] respectively, p = .20). PG was associated with a significantly lower incidence of major or minor vascular complications, compared to PS (adjusted OR: 0.54; 95% CI: 0.37-0.80; p < .01). Rates of acute kidney injury were also lower with the PG device. There was no significant difference between bleeding, MACCE, and death.
CONCLUSIONS
Compared to PS, the PG VCD was associated with a lower rate of major or minor vascular complications and lower rates of AKI after transfemoral TAVR.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.