MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tbme.2023.3293952

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2023

Portal Vein Pressure Estimation and Portal Hypertension Discrimination Based on Subharmonic Scattering of Ultrasound Contrast Agent Microbubbles.

Xu Gang, Wang Yun, Lu Huimin, Li Changcan, Huang Laixin, Su Min, Yang Huayu, Li Deyu, Qiu Weibao, Sun Minghao, Yang Jiayin, Mao Yilei, Li Fei

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

The acquisition of real-time portal vein pressure (PVP) is important for portal hypertension (PH) discrimination to monitor disease progress and select treatment options. To date, the PVP evaluation approaches are either invasive or noninvasive but with less stability and sensitivity.

METHODS

We customized an open ultrasound scanner to explore in vitro and in vivo the ultrasound contrast agent SonoVue microbubbles' subharmonic characteristics with acoustic pressure and local ambient pressure, and obtained promising results of PVP measurements in canine models with induced PH by ligation or embolization of portal vein.

RESULTS

In in vitro experiments, the highest correlations between the subharmonic amplitude of SonoVue microbubbles and ambient pressure were observed at acoustic pressures of 523 kPa and 563 kPa (r = -0.993, -0.993, P<0.05, respectively). The correlation coefficients between absolute subharmonic amplitudes and PVP (10.7-35.4 mmHg) were the highest among existing studies using microbubbles as pressure sensors (r values ranged from -0.819 to -0.918). The PH (>16 mmHg) diagnostic capacity also achieved a high level (563 kPa, sensitivity = 93.3%, specificity = 91.7%, accuracy = 92.6%).

CONCLUSION

This study proposes a promising measurement for PVP with the highest accuracy, sensitivity, and specificity in an in vivo model compared to existing studies. Future investigations are planned to assess the feasibility of this technique in clinical practice.

SIGNIFICANCE: This is the first study that comprehensively investigates the role of the subharmonic scattering signals from SonoVue microbubbles in evaluating PVP in vivo. It represents a promising alternative to invasive measurements for portal pressure.

Для цитирования: Xu Gang, Wang Yun, Lu Huimin, Li Changcan, Huang Laixin, Su Min, Yang Huayu, Li Deyu, Qiu Weibao, Sun Minghao, Yang Jiayin, Mao Yilei, Li Fei. Portal Vein Pressure Estimation and Portal Hypertension Discrimination Based on Subharmonic Scattering of Ultrasound Contrast Agent Microbubbles.. IEEE transactions on bio-medical engineering. 2023. DOI: 10.1109/tbme.2023.3293952.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.