PURPOSE: The primary stability of dental implants is one of the most crucial factors for providing long-term success of osseointegration. Vertical deficiencies, such as those due to maxillary sinus pneumatization, may cause a severe vertical limitation to residual bone height. This study aimed to examine the primary stabilization of implants without apical contacts.

MATERIALS AND

METHODS

Eighty bone-level implants (4.1-mm diameter/10-mm length) were placed into polyurethane test blocks without apical contacts. According to coronal bone-to-implant contact, groups were set as 4, 6, 8, and 10 mm, respectively. Resonance frequency analysis (RFA) using a SmartPeg was performed separately toward the transversal and horizontal axes by two independent researchers. Data were statistically compared for interobserver and among groups.

RESULTS

Interobserver reliability varied from moderate to excellent (intraclass correlation coefficient [ICC]: 0.629 to 0.985). There were no significant differences among the 6 mm, 8 mm, and 10 mm groups, although the 4 mm group showed the significantly lowest stability (P < .001). Transversal and longitudinal measurements of the same groups did not show a parallel correlation statistically.

CONCLUSION

RFA values may be affected by the finger torque in tightening of the SmartPeg among different researchers. Fully placed implants did not significantly show the highest stability among various apically contactless groups. Consequently, RFA should not be used alone to evaluate primary stability for implants without an apical contact.

Для цитирования: Özyurt Anıl, Bilmenoğlu Çağlar. Is Resonance Frequency Analysis a Reliable Evaluation for Primary Stability of Implants Without Apical Contact?. The International journal of oral & maxillofacial implants. 2022. DOI: 10.11607/jomi.9480.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.