MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/lary.27418

Медицинская библиотека · The Laryngoscope · 2018

Measurement reliability of phonation threshold pressure in pediatric subjects.

Hoffman Matthew R, Scholp Austin J, Hedberg Calvin D, Lamb Jim R, Braden Maia N, McMurray J Scott, Jiang Jack J

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES/HYPOTHESIS: Phonation threshold pressure (PTP), the minimum subglottal pressure (Ps ) required for phonation, is sensitive to changes in laryngeal biomechanics and is often elevated with pathology. Little is reported on PTP in children; challenges with task performance and measurement reliability represent barriers to routine clinical assessment.

STUDY

DESIGN

Pilot study evaluating PTP and Ps measurement reliability in children using labial and mechanical interruption.

METHODS

Twenty-two subjects aged 4 to 17 years (10.7 ± 3.9 years) participated. Ten trials were performed for each method; task order was randomized. For labial interruption, subjects produced /pα/ five times at softest (onset PTP) and comfortable amplitude. For mechanical interruption, subjects produced a sustained /α/ while a balloon valve interrupted phonation five times for 250 ms each; mechanical interruption was performed with a mouthpiece and mask. PTP was recorded as the difference between Ps and supraglottal pressure at phonation cessation (offset PTP). Mean PTP and Ps and intrasubject coefficients of variation were compared. Correlations with age were evaluated.

RESULTS

Mean PTP (P < .001) and Ps (P = .005) were higher for labial interruption. Intrasubject coefficients of variation for PTP (P = .554) and Ps (P = .305) were similar across methods. Coefficient of variation was related to age for mechanical-mask trials only (r = -0.628, P = .00175).

CONCLUSIONS

Differences in means are likely related to differences in task and PTP hysteresis effect. Reliability is comparable with all methods; using a mouthpiece may be preferable to a mask for mechanical interruption. Measurement of PTP is noninvasive, reliable, and may be a useful adjunct in pediatric voice assessment.

LEVEL OF EVIDENCE: 3b Laryngoscope, 129:1520-1526, 2019.

Для цитирования: Hoffman Matthew R, Scholp Austin J, Hedberg Calvin D, Lamb Jim R, Braden Maia N, McMurray J Scott, Jiang Jack J. Measurement reliability of phonation threshold pressure in pediatric subjects.. The Laryngoscope. 2018. DOI: 10.1002/lary.27418.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.