DOI 10.1007/s12350-021-02715-6

Медицинская библиотека · Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2021

An optimized imaging protocol for [99mTc]Tc-DPD scintigraphy and SPECT/CT quantification in cardiac transthyretin (ATTR) amyloidosis.

Schatka Imke, Bingel Anne, Schau Franziska, Bluemel Stephanie, Messroghli Daniel R, Frumkin David, Knebel Fabian, Diekmann Sonja M, Elsanhoury Ahmed, Tschöpe Carsten, Hahn Katrin, Amthauer Holger, Rogasch Julian M M, Wetz Christoph

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

In [99mTc]Tc-DPD scintigraphy for myocardial ATTR amyloidosis, planar images 3 hour p.i. and SPECT/CT acquisition in L-mode are recommended. This study investigated if earlier planar images (1 hour p.i.) are beneficial and if SPECT/CT acquisition should be preferred in H-mode (180° detector angle) or L-mode (90°).

METHODS

In SPECT/CT phantom measurements (NaI cameras, N = 2; CZT, N = 1), peak contrast recovery (CRpeak) was derived from sphere inserts or myocardial insert (cardiac phantom; signal-to-background ratio [SBR], 10:1 or 5:1). In 25 positive and 38 negative patients (reference: endomyocardial biopsy or clinical diagnosis), Perugini scores and heart-to-contralateral (H/CL) count ratios were derived from planar images 1 hour and 3 hour p.i.

RESULTS

In phantom measurements, accuracy of myocardial CRpeak at SBR 10:1 (H-mode, 0.95-0.99) and reproducibility at 5:1 (H-mode, 1.02-1.14) was comparable for H-mode and L-mode. However, L-mode showed higher variability of background counts and sphere CRpeak throughout the field of view than H-mode. In patients, sensitivity/specificity were ≥ 95% for H/CL ratios at both time points and visual scoring 3 hour. At 1 hour, visual scores showed specificity of 89% and reduced reader's confidence.

CONCLUSIONS

Early DPD images provided no additional value for visual scoring or H/CL ratios. In SPECT/CT, H-mode is preferred over L-mode, especially if quantification is applied apart from the myocardium.

Для цитирования: Schatka Imke, Bingel Anne, Schau Franziska, Bluemel Stephanie, Messroghli Daniel R, Frumkin David, Knebel Fabian, Diekmann Sonja M, Elsanhoury Ahmed, Tschöpe Carsten, Hahn Katrin, Amthauer Holger, Rogasch Julian M M, Wetz Christoph. An optimized imaging protocol for [99mTc]Tc-DPD scintigraphy and SPECT/CT quantification in cardiac transthyretin (ATTR) amyloidosis.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2021. DOI: 10.1007/s12350-021-02715-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.