Prospective Midterm Outcomes of the PULSTA Self-Expandable Transcatheter Pulmonary Valve: The PULSTA CE Approval Study.

Lee Sang-Yun, Kim Gi Beom, Carminati Mario, Giugno Luca, Çelebi Ahmet, Yucel Ilker Kemal, Eicken Andreas, Ewert Peter, Georgiev Stanimir, Choi Jae Young, Kim Chang Sin, Gutiérrez-Larraya Federico, Balbacid Domingo Enrique, Odemis Ender, Kizilkaya Mete Han, Luis Zunzunegui José, Ballesteros Tejerizo Fernando, Kim Seong-Ho, Jang So-Ick, Krasemann Thomas, Krings Gregor

СтатьяМногоцентровое исследованиеКлиническое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Patients with congenital or acquired right ventricular outflow tract dysfunction frequently require pulmonary valve replacement and are exposed to repeated surgical interventions due to prosthetic valve degeneration. The PULSTA self-expanding transcatheter pulmonary valve was designed to address a broad spectrum of native and surgically repaired right ventricular outflow tract anatomies. This study reports the midterm safety, hemodynamic performance, and durability outcomes of the prospective, multinational PULSTA CE approval trial (PULSTA Carpentier Edwards).

METHODS

This premarket, multicenter, nonrandomized study enrolled 58 patients across 11 centers in 6 countries. Successful PULSTA implantation was achieved in 57 patients (98.3%), including 4 valve-in-valve procedures. Patients were followed for a median duration of 4.1 years. Primary end points included procedural or device-related serious adverse events, hemodynamic improvement, and sustained valve function.

RESULTS

At 6 months, cardiac magnetic resonance imaging demonstrated significant reverse right ventricular remodeling, with marked reductions in right ventricular end-diastolic volume index, end-systolic volume index, and pulmonary regurgitation fraction (all P<0.001). Valve performance remained stable throughout follow-up, with 93.8% of patients exhibiting mild or less pulmonary regurgitation at 4.1 years. Freedom from reintervention was 98.2% at 5 years. One patient required a PULSTA-in-PULSTA implantation due to paravalvular leakage, and one case of infective endocarditis was successfully managed with antibiotic therapy. No stent fractures or structural valve deterioration were observed.

CONCLUSIONS

The PULSTA transcatheter pulmonary valve demonstrated high procedural success, durable valve function, and sustained hemodynamic benefits. Its adaptability to diverse right ventricular outflow tract anatomies and favorable safety profile supports its role as an effective therapeutic option for patients with right ventricular outflow tract dysfunction.

REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT03983512.

Для цитирования: Lee Sang-Yun, Kim Gi Beom, Carminati Mario, Giugno Luca, Çelebi Ahmet, Yucel Ilker Kemal, Eicken Andreas, Ewert Peter, Georgiev Stanimir, Choi Jae Young, Kim Chang Sin, Gutiérrez-Larraya Federico, Balbacid Domingo Enrique, Odemis Ender, Kizilkaya Mete Han, Luis Zunzunegui José, Ballesteros Tejerizo Fernando, Kim Seong-Ho, Jang So-Ick, Krasemann Thomas, Krings Gregor. Prospective Midterm Outcomes of the PULSTA Self-Expandable Transcatheter Pulmonary Valve: The PULSTA CE Approval Study.. Circulation. Cardiovascular interventions. 2026. DOI: 10.1161/circinterventions.126.016616.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.