How does ankle power on the prosthetic side influence loading parameters on the sound side during level walking of persons with transfemoral amputation?
Аннотация
BACKGROUND
Increased ankle power on the prosthetic side seems to decrease biomechanical loading parameters on the sound side. This assumption is based on biomechanical comparisons of different foot constructions. However, such study designs could not show whether the amount of ankle power solely influences the sound side.
OBJECTIVE
To analyze the influence of divergent ankle power, resulting from different foot constructions and from different ankle power settings, on the sound side loading parameters.
STUDY
DESIGN
Interventional cross sectional study.
METHODS
Level walking of transfemoral amputees with a microprocessor knee joint and Solid Ankle Cushioned Heel (SACH), energy storing and returning (ESR) and powered foot (PF) was analyzed. The PF was adapted in three configurations: without power (np), low power (lp), and optimal power (op). An optoelectronic camera system with 12 cameras and two force plates were used.
RESULTS
The ankle power on the prosthetic side shows significant differences about foot types and different settings of the PF. The knee adduction moment, the knee flexion moment, and the vertical ground reaction forces on the sound side were significantly reduced with PF_op and ESR in comparison to SACH. When analyzing these parameters for the different PF configurations, only some show significant results at normal velocity.
CONCLUSIONS
The additional positive mechanical work for an active push off in the PF tends to have a relieving effect. The biomechanical sound side loading parameters are reduced with PF_op in comparison to SACH and ESR, resulting in a relief of the sound side of lower limb amputees.