Cone beam CT (CBCT) in radiotherapy: Assessment of doses using a pragmatic setup in an international setting.

Djukelic Mario, Martin Colin John, Abuhaimed Abdullah, Kron Tomas, Gros Sebastien, Wood Tim, Pankowski Piotr, Ung Ngie Min, Vassileva Jenia, Plazas María Cristina, Vostinic Snezana, Lazovic Anja, Sá Ana Cravo, Nilsson Isabelle, Koutrouli Marianna, Murugan Lavanya, Fourie Hein, Miadzvetski Aliaksandr, Ameri Buthaina Al, Dumancic Mirta, Cheung Anson Ho-Yin, Rahbi Zakiya Al, Roussakis Yiannis, Shaaban Hossam Ragab, Liang Runcheng, Tomic Nada, Salazar Correa Daniel Eduardo, Butler Duncan, Small William

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The imaging modality kV CBCT on linear accelerators (linacs) is utilised to verify positioning and anatomy in cancer patients undergoing radiotherapy treatment. There is a need for optimisation of radiological protection in kV CBCT imaging protocols to avoid unnecessarily high exposures to normal tissues surrounding the target.

METHODS

A network of ICRP mentees from 23 countries were surveyed for available dosimetry equipment. Standardised measurements on CBCT linac imaging systems were conducted using a cone beam dose index (CBDI) devised as a straightforward measurement for wide beam doses. Measurements were made with (a) 100 mm ionisation chambers or (b) 0.6 cc Farmer ionisation chambers and cylindrical CT PMMA phantoms, and (c) an alternative setup of Farmer chambers and cubical phantoms comprised of slabs of water equivalent material readily available in radiotherapy centres. The measurements were compared with Monte Carlo (MC) simulations.

RESULTS

The survey showed limited availability for the reference setup using 100 mm chambers and CT phantoms. Correction factors were derived to convert normalised CBDI from alternative setups to the reference setup and are on average within 2% of MC simulations.

CONCLUSION

The slab phantom in combination with a Farmer chamber provides an alternative to quantify CBCT radiation dose indices from linac-based image-guided radiotherapy using materials accessible in most centres worldwide. A method is presented to use correction factors for Varian Truebeam linacs if traditional 100 mm chambers and cylindrical CT phantoms are not available. This will enable most radiotherapy centres across the world to engage in meaningful imaging dose measurement and optimisation.

Для цитирования: Djukelic Mario, Martin Colin John, Abuhaimed Abdullah, Kron Tomas, Gros Sebastien, Wood Tim, Pankowski Piotr, Ung Ngie Min, Vassileva Jenia, Plazas María Cristina, Vostinic Snezana, Lazovic Anja, Sá Ana Cravo, Nilsson Isabelle, Koutrouli Marianna, Murugan Lavanya, Fourie Hein, Miadzvetski Aliaksandr, Ameri Buthaina Al, Dumancic Mirta, Cheung Anson Ho-Yin, Rahbi Zakiya Al, Roussakis Yiannis, Shaaban Hossam Ragab, Liang Runcheng, Tomic Nada, Salazar Correa Daniel Eduardo, Butler Duncan, Small William. Cone beam CT (CBCT) in radiotherapy: Assessment of doses using a pragmatic setup in an international setting.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2025. DOI: 10.1016/j.ejmp.2025.104937.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.