MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/den.70028

Медицинская библиотека · Digestive endoscopy : official journal of the Japan Gastroenterological Endoscopy Society · 2025

Vendor-Agnostic Vision Transformer-Based Artificial Intelligence for Peroral Cholangioscopy: Diagnostic Performance in Biliary Strictures Compared With Convolutional Neural Networks and Endoscopists.

Sato Ryosuke, Matsumoto Kazuyuki, Tomiya Masahiro, Tanimoto Takayoshi, Ohto Akimitsu, Oki Kentaro, Kajitani Satoshi, Kikuchi Tatsuya, Matsumi Akihiro, Miyamoto Kazuya, Fujii Yuki, Uchida Daisuke, Tsutsumi Koichiro, Horiguchi Shigeru, Kawahara Yoshiro, Otsuka Motoyuki

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Accurate diagnosis of biliary strictures remains challenging. This study aimed to develop an artificial intelligence (AI) system for peroral cholangioscopy (POCS) using a Vision Transformer (ViT) architecture and to evaluate its performance compared to different vendor devices, conventional convolutional neural networks (CNNs), and endoscopists.

METHODS

We retrospectively analyzed 125 patients with indeterminate biliary strictures who underwent POCS between 2012 and 2024. AI models including the ViT architecture and two established CNN architectures were developed using images from CHF-B260 or B290 (CHF group; Olympus Medical) and SpyScope DS or DS II (Spy group; Boston Scientific) systems via a patient-level, 3-fold cross-validation. For a direct comparison against endoscopists, a balanced 440-image test set, containing an equal number of images from each vendor, was used for a blinded evaluation.

RESULTS

The 3-fold cross-validation on the entire 2062-image dataset yielded a robust accuracy of 83.9% (95% confidence interval (CI), 80.9-86.7) for the ViT model. The model's accuracy was consistent between CHF (82.7%) and Spy (86.8%, p = 0.198) groups, and its performance was comparable to the evaluated conventional CNNs. On the 440-image test set, the ViT's accuracy of 78.4% (95% CI, 72.5-83.8) was comparable to that of expert endoscopists (82.0%, p = 0.148) and non-experts (73.0%, p = 0.066), with no statistically significant differences observed.

CONCLUSIONS

The novel ViT-based AI model demonstrated high vendor-agnostic diagnostic accuracy across multiple POCS systems, achieving performance comparable to conventional CNNs and endoscopists evaluated in this study.

Для цитирования: Sato Ryosuke, Matsumoto Kazuyuki, Tomiya Masahiro, Tanimoto Takayoshi, Ohto Akimitsu, Oki Kentaro, Kajitani Satoshi, Kikuchi Tatsuya, Matsumi Akihiro, Miyamoto Kazuya, Fujii Yuki, Uchida Daisuke, Tsutsumi Koichiro, Horiguchi Shigeru, Kawahara Yoshiro, Otsuka Motoyuki. Vendor-Agnostic Vision Transformer-Based Artificial Intelligence for Peroral Cholangioscopy: Diagnostic Performance in Biliary Strictures Compared With Convolutional Neural Networks and Endoscopists.. Digestive endoscopy : official journal of the Japan Gastroenterological Endoscopy Society. 2025. DOI: 10.1111/den.70028.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.