MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.neuroimage.2009.04.044

Медицинская библиотека · NeuroImage · 2009

Safety of MRI in patients with implanted deep brain stimulation devices.

Tagliati Michele, Jankovic Joseph, Pagan Fernando, Susatia Frandy, Isaias Ioannis U, Okun Michael S, National Parkinson Foundation DBS Working Group

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To survey the safety of MRI in PD patients implanted with DBS devices.

BACKGROUND

MRI in patients with DBS implants is useful to confirm electrode placement, optimize programming and investigating complications. However, several medical centers do not perform MRI studies in DBS patients because of safety concerns. The safety profile of MRI in DBS patients has not been documented in large clinical series.

METHODS

42 NPF Centers of Excellence (COEs) were asked to complete a questionnaire on MRI use and DBS.

RESULTS

Investigators from 40 of 42 (95%) NPF COEs completed the survey and 23 (58%) reported that they were currently performing brain MRI in DBS patients, while 3 (7.5%) had done it in the past. The 17 COEs currently not performing post-operative MRI for DBS listed the following reasons: 1) industry guidelines and/or warnings (53%); 2) decision deferred to outside department (29%); 3) liability/risk/safety (18%); 4) no active DBS program (18%); 5) no available MRI (12%); and 6) insurance and reimbursement concerns (6%). A total of 3304 PD patients with one or more DBS leads had a brain MRI scan, and 177 DBS patients had MRI of other body regions. In one case MRI was associated with an IPG failure without neurological sequelae after IPG replacement. No other complications were reported.

CONCLUSIONS

these data provide evidence for a favorable risk/benefit ratio for brain MRI in patients with DBS implants. Further studies will need to address whether a re-assessment of more restrictive recommendations (i.e. very low SAR values) may be warranted.

Для цитирования: Tagliati Michele, Jankovic Joseph, Pagan Fernando, Susatia Frandy, Isaias Ioannis U, Okun Michael S, National Parkinson Foundation DBS Working Group. Safety of MRI in patients with implanted deep brain stimulation devices.. NeuroImage. 2009. DOI: 10.1016/j.neuroimage.2009.04.044.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.