Performance of PROPELLER relative to standard FSE T2-weighted imaging in pediatric brain MRI.

Vertinsky A Talia, Rubesova Erika, Krasnokutsky Michael V, Bammer Sabine, Rosenberg Jarrett, White Allan, Barnes Patrick D, Bammer Roland

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

T2-weighted fast spin-echo imaging (T2-W FSE) is frequently degraded by motion in pediatric patients. MR imaging with periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) employs alternate sampling of k-space to achieve motion reduction.

OBJECTIVE

To compare T2-W PROPELLER FSE (T2-W PROP) with conventional T2-W FSE for: (1) image quality; (2) presence of artefacts; and (3) ability to detect lesions.

MATERIALS AND

METHODS

Ninety-five pediatric patients undergoing brain MRI (1.5 T) were evaluated with T2-W FSE and T2-W PROP. Three independent radiologists rated T2-W FSE and T2-W PROP, assessing image quality, presence of artefacts, and diagnostic confidence. Chi-square analysis and Wilcoxon signed rank test were used to assess the radiologists' responses.

RESULTS

Compared with T2-W FSE, T2-W PROP demonstrated better image quality and reduced motion artefacts, with the greatest benefit in children younger than 6 months. Although detection rates were comparable for the two sequences, blood products were more conspicuous on T2-W FSE. Diagnostic confidence was higher using T2-W PROP in children younger than 6 months. Average inter-rater agreement was 87%.

CONCLUSION

T2-W PROP showed reduced motion artefacts and improved diagnostic confidence in children younger than 6 months. Thus, use of T2-W PROP rather than T2-W FSE should be considered in routine imaging of this age group, with caution required in identifying blood products.

Для цитирования: Vertinsky A Talia, Rubesova Erika, Krasnokutsky Michael V, Bammer Sabine, Rosenberg Jarrett, White Allan, Barnes Patrick D, Bammer Roland. Performance of PROPELLER relative to standard FSE T2-weighted imaging in pediatric brain MRI.. Pediatric radiology. 2009. DOI: 10.1007/s00247-009-1292-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.