DOI 10.1111/dom.14609

Медицинская библиотека · Diabetes, obesity & metabolism · 2021

Performance of intermittently scanned continuous glucose monitoring systems in people with type 1 diabetes: A pooled analysis.

Moser Othmar, Sternad Christoph, Eckstein Max L, Szadkowska Agnieszka, Michalak Arkadiusz, Mader Julia K, Ziko Haris, Elsayed Hesham, Aberer Felix, Sola-Gazagnes Agnes, Larger Etienne, Fadini Gian Poalo, Bonora Benedetta Maria, Bruttomesso Daniela, Boscari Federico, Freckmann Guido, Pleus Stefan, Christiansen Sverre C, Sourij Harald

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: To conduct a pooled analysis to assess the performance of intermittently scanned continuous glucose monitoring (isCGM) in association with the rate of change in sensor glucose in a cohort of children, adolescents, and adults with type 1 diabetes.

MATERIAL AND

METHODS

In this pooled analysis, isCGM system accuracy was assessed depending on the rate of change in sensor glucose. Clinical studies that have been investigating isCGM accuracy against blood glucose, accompanied with collection time points were included in this analysis. isCGM performance was assessed by means of median absolute relative difference (MedARD), Parkes error grid (PEG) and Bland-Altman plot analyses.

RESULTS

Twelve studies comprising 311 participants were included, with a total of 15 837 paired measurements. The overall MedARD (interquartile range) was 12.7% (5.9-23.5) and MedARD differed significantly based on the rate of change in glucose (P < 0.001). An absolute difference of -22 mg/dL (-1.2 mmol/L) (95% limits of agreement [LoA] 60 mg/dL (3.3 mmol/L), -103 mg/dL (-5.7 mmol/L)) was found when glucose was rapidly increasing (isCGM glucose minus reference blood glucose), while a -32 mg/dL (1.8 mmol/L) (95% LoA 116 mg/dL (6.4 mmol/L), -51 mg/dL (-2.8 mmol/L)) absolute difference was observed in periods of rapidly decreasing glucose.

CONCLUSIONS

The performance of isCGM was good when compared to reference blood glucose measurements. The rate of change in glucose for both increasing and decreasing glucose levels diminished isCGM performance, showing lower accuracy during high rates of glucose change.

Для цитирования: Moser Othmar, Sternad Christoph, Eckstein Max L, Szadkowska Agnieszka, Michalak Arkadiusz, Mader Julia K, Ziko Haris, Elsayed Hesham, Aberer Felix, Sola-Gazagnes Agnes, Larger Etienne, Fadini Gian Poalo, Bonora Benedetta Maria, Bruttomesso Daniela, Boscari Federico, Freckmann Guido, Pleus Stefan, Christiansen Sverre C, Sourij Harald. Performance of intermittently scanned continuous glucose monitoring systems in people with type 1 diabetes: A pooled analysis.. Diabetes, obesity & metabolism. 2021. DOI: 10.1111/dom.14609.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.