MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tbme.2008.919715

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2008

Accelerometers and force sensing resistors for optimal control of walking of a hemiplegic.

Došen Strahinja, Popovi Dejan B

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

We developed a method for use of accelerometers and force sensing resistors (FSRs) within an optimal controller of walking for hemiplegic individuals. The data from four dual-axis accelerometers and four FSRs were inputs, while six muscle activation profiles were outputs. The controller includes two stages: 1) estimating the target gait pattern using artificial neural networks; and 2) optimal control minimizing tracking errors (from the estimated gait pattern) and muscle efforts. The controller was tested using data collected from six healthy subjects walking at five speeds (0.6-1.4 m/s). The average root mean square errors (RMSEs) normalized by the peak-to-peak value of the target signals [normalized RMSE (NRMSE)] were below 6%, 7%, 8%, and 3% for estimation of joint angles, hip acceleration, ground reaction force, and movement of the center of pressure, respectively. Using the estimated data as inputs, the simulation generated the target healthy-like gait patterns and reproducible muscle activation profiles in 90% of 300 tested gait trials. Overall tracking NRMSE was between 2% and 9%. The optimal controller was developed for testing the feasibility of healthy-like gait patterns in hemiplegic individuals, and generating a knowledge base that is required for the synthesis of a sensory-driven control of walking assisted by functional electrical stimulation.

Для цитирования: Došen Strahinja, Popovi Dejan B. Accelerometers and force sensing resistors for optimal control of walking of a hemiplegic.. IEEE transactions on bio-medical engineering. 2008. DOI: 10.1109/tbme.2008.919715.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.