Accuracy Evaluation of Collapsed Cone Convolution Superposition Algorithms for the Nasopharynx Interface in the Early Stage of Nasopharyngeal Carcinoma.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lai Yuan-Chun, Hung Li-Chung, Jeng Chien-Chung, Feng Chen-Ju, Chang Tung-Hao, Hsu Shih-Ming
Аннотация
This study combined the use of radiation dosimeteric measurements and a custom-made anthropomorphic phantom in order to evaluate the accuracy of therapeutic dose calculations at the nasopharyngeal air-tissue interface. The doses at the nasopharyngeal air-tissue interface obtained utilizing the Pinnacle and TomoTherapy TPS, which are based on collapsed cone convolution superposition (CCCS) algorithms, were evaluated and measured under single 10 × 10 cm2, 2 × 2 cm2, two parallel opposed 2 × 2 cm2 and clinical fields for early stage of nasopharyngeal carcinoma by using EBT3, GR-200F, and TLD 100. At the air-tissue interface under a 10 × 10 cm2 field, the TPS dose calculation values were in good agreement with the dosimeter measurement with all differences within 3.5%. When measured the single field 2 × 2 cm2, the differences between the average dose were measured at the distal interface for EBT3, GR-200F, and TLD-100 and the calculation values were -15.8%, -16.4%, and -4.9%, respectively. When using the clinical techniques such as IMRT, VMAT, and tomotherapy, the measurement results at the interface for all three techniques did not imply under dose. Small-field sizes will lead to dose overestimation at the nasopharyngeal air-tissue interface due to electronic disequilibrium when using CCCS algorithms. However, under clinical applications of multiangle irradiation, the dose errors caused by this effect were not significant.