Impulse oscillometry bronchodilator response in preschool children.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Meoli Aniello, Trischler Jordis, Hutter Martin, Dressler Melanie, Esposito Susanna, Blümchen Katharina, Zielen Stefan, Schulze Johannes

Аннотация

BACKGROUND: In preschoolers, performing an acceptable spirometry and measuring bronchodilator response (BDR) is challenging; in this context, impulse oscillometry (IOS) represents a valid alternative. However, more studies on the standardization of BDR for IOS in young children are required.

OBJECTIVE: The objective of the study was to identify optimal thresholds to define a positive BDR test with IOS in preschoolers with suspected asthma.

METHODS: Children aged 3-6 years with suspected asthma and their lung function investigated with both IOS and spirometry pre- and post-BDR were retrospectively analyzed. The spirometric BDR was defined as positive when the change of FEV1 was ≥12% or ≥200 mL. The oscillometric BDR was defined as positive in case of change of at least -40% in R5, +50% in X5, and -80% in AX.

RESULTS: Among 72 patients, 36 (age 5.2 ± 1 years; 64% boys) were selected for the subsequent analysis according to ATS/ERS quality criteria of measurements; specifically, 19 patients did not meet IOS and 36 did not meet spirometry criteria. The spirometric BDR was found positive in seven subjects (19.4%); conversely, a positive oscillometric BDR was identified in four patients (11.1%). No patient presented a positive BDR response with both methods. In IOS, the mean decrease in R5 and AX was 19.9% ± 10% and 44% ± 22.1%, and the mean increase in X5 was 23.3% ± 17.8%, respectively. A decrease in R5 of 25.7% (AUC 0.77, p = .03) and an increase in X5 of 25.7% (AUC 0.75, p = .04) showed the best combination of sensitivity and specificity to detect an increase of FEV1 ≥ 12% and/or ≥200 mL.

CONCLUSION: The IOS represents a valid alternative to spirometry to measure BDR in preschool children and should be the gold standard in this age group. We are considering a decrease of 26% in R5 and an increase of 26% in X5 as diagnostic threshold for BDR.