Does Meal or Water Intake Affect Ultrasound Attenuation Coefficient Estimate?
Аннотация
OBJECTIVES
To assess whether meal or water intake may affect the measurement of the ultrasound (US) attenuation coefficient (AC) imaging, a parameter that is directly related to liver fat content.
METHODS
The study was performed in two centers (Italy and USA). AC was obtained using the ATI algorithm implemented in the Aplio i-series US systems (Canon Medical Systems, Japan) by one operator at each center. Measurements were performed at baseline and 5, 15, 30, 45 minutes after drinking 500 mL of water (group 1), or 30, 45, 60, 90, 120 minutes after eating a meal of about 600 kcal (group 2). Multilevel generalized estimating equations for repeated measures were used for the statistical analysis to consider the clustered nature of the data.
RESULTS
Twenty-six individuals were enrolled: 11 (10 females; age, 43.7 ± 12.5 years) in Italy and 15 (10 females; age, 60.7 ± 6.3 years) in USA. At B-mode US, 10 (38.5%) had liver steatosis. The baseline AC values, in decibel/centimeter/megahertz, were 0.64 (0.12) in group 1 and 0.66 (0.13) in group 2. There was not any significant difference in AC values at every time-point after water or meal intake either in group 1 or group 2. This result did not change including sex, age, and skin-to-liver capsule into the models.
CONCLUSIONS
The measurement of the AC, which is a biomarker of liver steatosis, does not require a fasting state and drinking water does not affect the result.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.