DOI 10.23736/s0021-9509.17.09305-3

Медицинская библиотека · The Journal of cardiovascular surgery · 2017

The INNOVATION Trial: four-year safety and effectiveness of the INCRAFT® AAA Stent-Graft System for endovascular repair.

Pratesi Giovanni, Pratesi Carlo, Chiesa Roberto, Coppi Gioacchino, Scheinert Dierk, Brunkwall Jan S, van der Meulen Stefaan, Torsello Giovanni

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

This paper reports the 4-year safety and effectiveness of the INCRAFT® AAA Stent-Graft System (Cordis Corp., Milpitas, CA, USA), an ultra-low-profile device for the treatment of abdominal aortic aneurysms.

METHODS

The INNOVATION Trial is the prospective, first-in-human, multicenter trial to evaluate the safety and effectiveness of the INCRAFT® System. Patients underwent annual clinical and computed tomography angiography examination as part of the study protocol. The INCRAFT® AAA Stent-Graft System is a customizable tri-modular design, with an ultra-low profile (14-Fr) delivery system. Patient were treated under approved protocol, the prescribed clinical and imaging follow-up at annually through 5 years. Results analyzed and adjudicated by a clinical events committee, independent core laboratory, and a data safety and monitoring board. This manuscript reports results through 4 years of follow-up.

RESULTS

A total of 60 patients were enrolled in the trial, all of whom were successfully treated. Follow-up rates at 1 and 4 years were 93% (56/60) and 85% (51/60), respectively. All-cause mortality at 4 years was 17.6% and no death was AAA-, device-, or procedure-related. The secondary reintervention rate at 1 year was 4.6%, primarily the result of stent thrombosis. In total, 10 patients required 13 post-procedure interventions within 4-years of follow-up (2 to repair a type I endoleak, 4 to repair a type II endoleak, 1 for stent thrombosis, 1 for renal stenosis, 1 for aneurysm enlargement, 2 for limb migration and 2 for prosthesis stenosis or occlusion). There were 4 cases (10%) of aneurysm enlargement reported at the 4 year follow-up. At 4 years, 38 out of 39 patients were free from type I and III endoleaks. There were no proximal type I or type III endoleaks at 4-year follow-up. Core laboratory evaluation of the postoperative imaging studies indicated absence of endograft migration while a single fracture was demonstrated without any clinical sequelae.

CONCLUSIONS

The INCRAFT® AAA Stent-Graft System provides a minimally invasive and durable solution for patients undergoing EVAR that has been associated with a low frequency of device-related events through 4 years of follow-up.

Для цитирования: Pratesi Giovanni, Pratesi Carlo, Chiesa Roberto, Coppi Gioacchino, Scheinert Dierk, Brunkwall Jan S, van der Meulen Stefaan, Torsello Giovanni. The INNOVATION Trial: four-year safety and effectiveness of the INCRAFT® AAA Stent-Graft System for endovascular repair.. The Journal of cardiovascular surgery. 2017. DOI: 10.23736/s0021-9509.17.09305-3.