Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise.

Nagayama Yasunori, Nakaura Takeshi, Tsuji Akinori, Urata Joji, Furusawa Mitsuhiro, Yuki Hideaki, Hirarta Kenichiro, Kidoh Masafumi, Oda Seitaro, Utsunomiya Daisuke, Yamashita Yasuyuki

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To retrospectively evaluate the image quality and radiation dose of 100-kVp scans with sinogram-affirmed iterative reconstruction (IR) for unenhanced head CT in adolescents.

METHODS

Sixty-nine patients aged 12-17 years underwent head CT under 120- (n = 34) or 100-kVp (n = 35) protocols. The 120-kVp images were reconstructed with filtered back-projection (FBP), 100-kVp images with FBP (100-kVp-F) and sinogram-affirmed IR (100-kVp-S). We compared the effective dose (ED), grey-white matter (GM-WM) contrast, image noise, and contrast-to-noise ratio (CNR) between protocols in supratentorial (ST) and posterior fossa (PS). We also assessed GM-WM contrast, image noise, sharpness, artifacts, and overall image quality on a four-point scale.

RESULTS

ED was 46% lower with 100- than 120-kVp (p < 0.001). GM-WM contrast was higher, and image noise was lower, on 100-kVp-S than 120-kVp at ST (p < 0.001). CNR of 100-kVp-S was higher than of 120-kVp (p < 0.001). GM-WM contrast of 100-kVp-S was subjectively rated as better than of 120-kVp (p < 0.001). There were no significant differences in the other criteria between 100-kVp-S and 120-kVp (p = 0.072-0.966).

CONCLUSIONS

The 100-kVp with sinogram-affirmed IR facilitated dramatic radiation reduction and better GM-WM contrast without increasing image noise in adolescent head CT.

KEY POINTS: • 100-kVp head CT provides 46% radiation dose reduction compared with 120-kVp. • 100-kVp scanning improves subjective and objective GM-WM contrast. • Sinogram-affirmed IR decreases head CT image noise, especially in supratentorial region. • 100-kVp protocol with sinogram-affirmed IR is suited for adolescent head CT.

Для цитирования: Nagayama Yasunori, Nakaura Takeshi, Tsuji Akinori, Urata Joji, Furusawa Mitsuhiro, Yuki Hideaki, Hirarta Kenichiro, Kidoh Masafumi, Oda Seitaro, Utsunomiya Daisuke, Yamashita Yasuyuki. Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise.. European radiology. 2016. DOI: 10.1007/s00330-016-4679-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.