MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/0003-4975(95)00757-c

Медицинская библиотека · The Annals of thoracic surgery · 1995

Ball valve (Smeloff-Cutter) aortic valve replacement without anticoagulation.

Gometza B, Duran C M

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Because of the difficulty of permanent anticoagulation in our young population, Smeloff-Cutter ball valves have been used since 1986 at our institution for aortic valve replacement in selected patients without permanent anticoagulation therapy.

METHODS

The availability of a satisfactory follow-up system since July 1988 suggested a study of all 47 patients operated on since then and followed for a mean of 43.2 months (range, 16 to 78 months). Mean age was 26.3 years, 98% were in sinus rhythm, and 16 patients (34%) had concomitant mitral repair.

RESULTS

There were no hospital deaths. Three patients were lost to follow-up at a mean of 27 months. Four late deaths occurred (8.5%), two of them sudden, with actuarial survival at 6 years of 91% +/- 4.3%. There were a total of five embolic events (2.9%/patient-year). For isolated aortic valve replacement only, with antiaggregant therapy (n = 29), the incidence was 0.9%/patient-year. For all patients receiving antiaggregant agents (n = 43), it was 3.02%/patient-year. There were no known cases of valve thrombosis. Reoperation was required in 5 patients.

CONCLUSION

Aortic valve replacement with the Smeloff-Cutter ball valve might be a valid alternative for young patients unable to maintain regular anticoagulation.

Для цитирования: Gometza B, Duran C M. Ball valve (Smeloff-Cutter) aortic valve replacement without anticoagulation.. The Annals of thoracic surgery. 1995. DOI: 10.1016/0003-4975(95)00757-c.