MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10840-016-0181-y

Медицинская библиотека · Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2016

Right atrial lead fixation type and lead position are associated with significant variation in complications.

Witt Chance M, Lenz Charles J, Shih Henry H, Ebrille Elisa, Rosenbaum Andrew N, Aung Htin, van Zyl Martin, Manocha Kevin K, Deshmukh Abhishek J, Hodge David O, Mulpuru Siva K, Cha Yong-Mei, Espinosa Raul E, Asirvatham Samuel J, McLeod Christopher J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Optimal atrial pacemaker lead position and fixation mechanism have not been determined with regard to effect on complications. We aimed to determine the association between atrial lead-related complications and varying atrial lead tip positions and lead fixation mechanisms.

METHODS

All patients who underwent dual-chamber pacemaker implant between 2004 and 2014 were retrospectively reviewed for atrial lead tip position and fixation type. Lead-related complications were assessed by electronic medical record review. Complication rates were compared at 1 year by chi-square analysis and at 5 years using a Kaplan-Meier analysis.

RESULTS

During the study period, 3451 patients (mean age 73.9, 53.4 % male) underwent dual-chamber pacemaker placement. Active fixation leads were associated with a higher incidence of pericardial effusion (81 (2.9 %) vs. 6 (1.0 %), p = 0.005) and pericardiocentesis (46 (1.6 %) vs. 2 (0.3 %), p = 0.01) at 1 year compared to passive fixation leads. There was no difference in overall complication rates by fixation type (161 (5.7 %) vs. 29 (4.6 %), p = 0.26). Low atrial septal lead tip position was associated with a higher rate of lead dislodgement (10 (15.2 %)) compared to appendage (46 (1.6 %)), free wall (10 (2.1 %)), or high atrial septal (2 (4.7 %)) positions (p < 0.001). This difference was also reflected in a significantly increased need for lead revision and overall complications. A multivariate analysis which included potential confounders confirmed the association of active fixation leads with an increased rate of perforation-related complications (p = 0.03) and septal lead location with increased rates of dislodgement (p < 0.001).

CONCLUSIONS

Active compared to passive lead fixation increases the risk for pericardial effusion requiring pericardiocentesis. There is a clear association between low atrial septal lead position and lead dislodgement requiring lead revision.

Для цитирования: Witt Chance M, Lenz Charles J, Shih Henry H, Ebrille Elisa, Rosenbaum Andrew N, Aung Htin, van Zyl Martin, Manocha Kevin K, Deshmukh Abhishek J, Hodge David O, Mulpuru Siva K, Cha Yong-Mei, Espinosa Raul E, Asirvatham Samuel J, McLeod Christopher J. Right atrial lead fixation type and lead position are associated with significant variation in complications.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2016. DOI: 10.1007/s10840-016-0181-y.